14
4.3.1 Verification of Calibration and Determination of Repair Costs
Flow chart for verifying the calibration of used tension meters, incoming and outgoing
verification with Inspection Certificate 3.1 according to DIN EN 10204
Delivery
Unit can be
calibrated
(functional)
Repair costs
exceed 40 % of as
new value
Costs approved
by customer
No
Yes
No
Yes
No
Yes
Vertification of the
calibration
Visual inspection
and functional test
Values within
specification
Prepare certification
for
incoming calibration
Determination
of costs
Make of
cost estimate
Repair and
adjustment
Verification of the
calibration
Calibration label
on the instrument
No
Invoice scrapping
and disposal
Invoice repair and
calibration services
and dispatch
Invoice
calibration services
and dispatch
Prepare certificate
for outgoing
calibration
Yes
Calibration label
on the instrument