Pilot²
Doc. No 609 2210 v04
9
Send worn or damaged contra-angles or contra-angles with material changes to the
Schlumbohm
®
customer service.
Processing cycles:
We recommend a regular service for the contra-angle after 1000 processing cycles or
one year.
Initial treatment at the point of use
•
Clean the medical device immediately after every treatment, to flush out any liquid
(e.g., blood, saliva etc.) and to prevent settling on the internal parts.
•
Operate the medical device for at least 10 seconds at idle speed
•
Ensure that all outlets are rinsed out
•
Wipe the entire surface of the instrument with disinfectant.
•
Remove the root canal instrument.
•
Remove the contra-angle from the motor
Note that the disinfectant used during pre-
treatment is only for personal protection and
cannot replace the disinfectant step after
cleaning.
EN