User manual SESD 216 / SESD 230
281606
page 16 of 20
7.3
Programming the automatic tests
Activate the setup function according chapter 8.1 and select a setup number, which you will program.
Principle it allows existing spaces to be overwritten. All 20 setup numbers SET 01 – SET 20 cannot
change, all contents of this can changed.
The 5 buttons and the potentiometer have double functions. In the setup mode they describe other things
Nr.
Normal Mode
Setup Mode
1
On / Off
ON / OFF switch
ON / OFF switch
2
Polarity
Polarity of the pulse,
positive and negative
Use
Use the adjusted
parameters
3
Con/Air
Mode CON / AIR
Contact discharge or air
discharge
Selection of various
parameters
4
Perm
Selection test level,
Continuous mode and
Setup - Mode
Escape
Back without saving the
adjusted parameters
5
Value
Potentiometer for
output voltage adjust
Value
Adjustment test level 1 – 4
and pre-selection parame-
ters
6
Counter
Counter pre-selectable
Clear
Clear of the setup values
7
Rep. Freq.
Repetition frequency
Store
Store of the setup values
8
LCD Display
4 lines
7.3.1
Input parameter
An example is for programming a blank space – e.g., SET 10. Right next to „SET 10“ is an arrow (left
arrow) to see. This arrow marks the position of the input fields. The arrow can be set using the key
„Con/Air“ at any position to move on. Values on the individual items can be set using the potentiometer.
Example:
-
MOD
The rotation of the potentiometers causes the indication AIR or CON. Similarly
some parameters are activated, depending on the choice AIR or CON.
-
REP
The repetition frequencies are stored and can be selected.
-
POL
Choice the polarity
-
SPA
The arrow points to the right, the „Starting voltage“ can be set.
-
STE
Enter the „voltage step“ or „voltage jump“.
-
END
Entering the „final voltage“ or the maximum test voltage.
-
NBSTEP
„Number of steps“ – indicating the number of test pulses per increment.
Depending on the voltage swing (beginning and end) and the step size results in
a maximum possible number of discharges per step. A higher number cannot be
set.