FRED PA-1
7.3 Defibrillation pulse
Form
• Biphasic truncated exponential waveform
• Maintains the energy delivered to the patient at an approximately constant level
with regard to patient resistance
ϰϬ
ϱϬ
ϲϬ
ϳϬ
ϴϬ
ϮϱKŚŵ
ϱϬKŚŵ
ϳϬKŚŵ
ͲϰϬ
ͲϯϬ
ͲϮϬ
ͲϭϬ
Ϭ
ϭϬ
ϮϬ
ϯϬ
ƵƌĂƚŝŽŶ;ŵƐ
Ϳ
ϭϬϬKŚŵ
ϭϮϱKŚŵ
ϭϱϬKŚŵ
ϮϮϬKŚŵ
ϮϱϬKŚŵ
Default energy settings
Accuracy at 50
Ω
: ± 3 J or ± 15 % (the higher value is assumed)
SCHILLER's customer service department can change the default energy levels to
the following values:
70 – 90 –120 – 150 – 200 J (adults)
30 – 50 – 70 J (children)
(automatic adaptation when paediatric pads are connected)
Cycle time: rhythm analysis –
shock availability (in semi-auto-
matic mode)
With full battery:
After 15 discharges with max.
energy:
(Maximum time between start of the analysis and shock availability, in semi-auto-
matic mode)
< 20 seconds
< 20 seconds
Patient impedance at which
shock delivery is possible
25 to 250
Ω
(Impedance is compensated up to 200
Ω
)
Indication when ready to shock
The orange button
is lit