Schaller SW1 Скачать руководство пользователя страница 5

version 1.1_n 

 

page 5 

Changing measuring interval: 

Menu – 

Options- Interval 

Standard is a measuring interval of 15 minutes. To increase the battery life, 
this interval can be exceeded to up to 12 hours. We recommend a measuring 
interval of about 15 minutes to one hour. 

 

 

Datalog:

 For activating the datalog, effect the following steps: 

 

Unlock:

 Menu - 

Options

 – 

Unlock 

The  standard  password  is  the  4-digit  serial  number.  Insert  the  number  and 
confirm by pressing OK. 

Changing user level

: Menu - 

Options

 – 

O User level 

For activating the advanced user, highlight this menu item.  

 

Datalog time: 

Menu – 

Options 

– 

Datalog time

 

Select your desired interval (except Hold) using the arrow keys and confirm 
by pressing 

OK

.  

Deactivating interval:  

Menu – 

Options 

– 

Interval 

Select „off” by pressing the   button and confirm by pressing 

OK

Starting the Autolog / saving measuring values 

Back in the measuring window press the   key to reach the store menu. For 
saving a measuring value or activating an automatic log, press the 

 button. 

For  completing  the  AutoLog,  switch  on  the  device  (if  necessary)  and  press 
the 

 button. If you want to add additional data please press the 

 button. 

Additional data can also be entered on the PC subsequently. 

 

For  returning  to  the  normal  use,  all  menu  items  have  to  be  reset:

 

Unlocking  by  entering  the  password,  changing  user  level  to  single  user  O, 
deactivating Autolog (or set to “manual”), changing the interval to 15 minutes, 
returning to the measuring window.  

 

Switching  on  the  display  lighting:

  Press  the 

  key  briefly;  the  display 

lighting switches off automatically after approx. 2 minutes. Pressing any key 
activates the display lighting again. 
 

Modifying display lighting time:

 Menu – 

Options 

– 

BL On Time

  

The  period  for  display  lighting  can  be  modified  by  pressing  the 

  and 

 

keys.

 

Auto  off:

 

The  instrument  switches  off  automatically  after  approx.  four 

minutes. 

 

Modifying Auto off time: 

Menu – 

Options 

– A

uto Off time 

The period for switching off can be modified by pressing the   and   keys.

 

 

Содержание SW1

Страница 1: ...User manual mould and air guard humimeter SW1 to prevent from health threats version 1 1_en Schaller GmbH 2017 ...

Страница 2: ...Preventing and avoiding mould Mount a wall hook on a position with increased potential of mould e g outer wall and position the humimeter SW1 device on the hook Switch on the instrument by pressing the button for 3 seconds Select the calibration curve Risk of mould using the arrow keys or The device switches on regularly and checks the level of risk In case of a high or very high risk of mould an ...

Страница 3: ...inition of calibration curves Risk of mould The calibration curve risk of mould clearly evaluates the risk grouped into 5 risk levels The risk levels are related directly to the relative air humidity calibration curve description unit measuring range risk of mould risk of mould No up to Very high abs humidity absolute air humidity g m 0 to 130 g m3 rel humidity relative air humidity rh 0 to 100 To...

Страница 4: ...s saturated with water vapour for half of its value 100 relative humidity means that the air is totally saturated When the air has more than 100 of relative humidity the excessive moisture would condense or form fog Operating the instrument Switching on Press for three seconds Changing the calibration curve or Menu Switch on the instrument pressing the key and immediately during showing the start ...

Страница 5: ...s the key to reach the store menu For saving a measuring value or activating an automatic log press the button For completing the AutoLog switch on the device if necessary and press the button If you want to add additional data please press the button Additional data can also be entered on the PC subsequently For returning to the normal use all menu items have to be reset Unlocking by entering the...

Страница 6: ...s Clear Logs Print Logs Last Logs All Logs Clear Logs Send Logs Manual Logs Auto Logs Clear Logs Options Date Time Interval Language Unlock C F DataLog Time Userlevel BL On Time Auto Off Time Materialcalib Password Reset Status Keypad symbols Measuring window Rolling menu Power ON OFF Switch upper Switch lower Save Hold Autolog Watch saved data Enter supplier s data Menu Enter Switch upper Switch ...

Страница 7: ...version 1 1_n page 7 Design of the device batteries slot flipside display of measuring value calibration curve ON button selection of calibration curve warning light wall holder ...

Страница 8: ...e measuring instrument and the material to measure resp the air Care instructions Do not drop the instrument or expose it to excessive temperatures The instrument is not waterproof Do not immerse the sensor in liquid The intervals for checking the instrument depend on your operational demands and the required level of accuracy In general the drift of the sensor according to the degree of use const...

Страница 9: ...tteries Press with your finger onto the arrow of the battery cap and pull it back Remove the empty batteries Put four new 1 5 Volt AA Alkaline batteries in the device Make sure that the position of the battery poles is correct Press down the batteries and close the cap If the battery symbol appears in the measuring window resp if a critical charge of battery is shown in the status the batteries ha...

Страница 10: ...ation automatic Data storage approx 10 000 measuring values Menu languages English German French Italian Spanish Russian Power supply 4 pcs of 1 5Volt AA Alkaline batteries Life time battery mould warning function approx 12 months ventilation monitor approx 1300 measurements Auto Off time after approx 4 minutes Power consumption 30 mA with display lighting Display 128 x 64 matrix display with LED ...

Страница 11: ...o too short conditioning time To demonstrate the importance of temperature adjustment the table below shows measuring errors due to a temperature difference of only 1 C 1 8 F between the measuring instrument and the substance to be measured at different ambient temperatures 10 C 50 F 20 C 68 F 30 C 86 F 10 r F 0 7 0 6 0 6 50 r F 3 5 3 2 3 0 90 r F 6 3 5 7 5 4 At room temperature 20 C 68 F and an a...

Страница 12: ...is blinking Ventilation monitor Calibration curve Absolute humidity If the measuring value INDOORS is HIGHER you can ventilate INDOORS OUTDOORS VENTILATE If the measuring value INDOORS is LOWER you may NOT ventilate Climate monitor Calibration curve relative humidity Ideal interior air humidity 40 60 A too high or too low value is felt as unpleasant and may cause health risks ...

Отзывы: