Version 2.4_en
page 3
Measuring procedure
1. For a correct measurement please ensure that the device has
the same temperature than the wood chips (+/-3°C). For that
reason, let your humimeter BLL adjust to the surrounding
temperature of the material for at least half an hour before
measuring.
2. Switch on the device: Press the
key
for 3 seconds.
3.
Change the calibration curve: Press one
time the
key
and then the
or key.
The name of the calibration curve can be
seen at the head of the display.
4. Plug the probe of your humimeter BLL
straight into the wood chips.
It is not
allowed to load the measuring head
incorrect or drop it down!
5.
Now the display shows the water
content.
Left hand the temperature of the
material is displayed.
6. To save the results in the save menu
press the ( button). The storage was
successful when the number in front of
the symbol
increased. To reach the
store menu please press ( ) until the
appears.
7. To name the saved results press the
button.
Risk of injury by measuring head! Keep away from children!