13
Remotefox that you would normally connect the PC
interface cable to and would read out the readings.
„Module 1“: The computer connection and thus also the connection for
the universal bus along with the alarm output of the first
data logger are now occupied with the connection line to
the Remotefox.
To be able to nevertheless connect a universal bus sensor
such as the "Temperature Multisensor Module", the
"Material Moisture Multisensor Module", the "Thermo-
Hygro-Sensor" etc. to the data logger, the Remotefox has
an additional jack named "Module 1".
This jack is used quasi as a "new" connection to the first
data logger and can be connected directly to the desired
universal bus sensors.
If the universal bus of the first data logger is not being
used, the "Module 1" jack serves as an alarm output for the
first measuring system.
„Logger 2“: The "Logger 2" input provides a facility for connecting a
second data logger to the Remotefox. The behaviour is the
same as for "Logger 1". The readings are sent in the same
rhythm and together with those of the first data logger per
email.
The
input
operates
otherwise
completely
independent of the "Logger 1" input.
„Module 2“: As in "Module 1", this jack serves quasi as a "new"
connection to the second data logger and can be connected
directly to the desired universal bus sensors, etc.