28
8.2 Setup of Limit Values by Actual Quantity Loads
1
Starting the limit function
Press and hold down the Set key.
Release the key when [L. SEt] is
displayed.
The currently set lower-limit value
flashes.
2
Loading the sample for the lower-limit value
Load the sample of the lower-limit value
on the measurement pan.
3
Saving the lower-limit value
Press the Memory key.
After the lower-limit value has been
saved, the balance displays it briefly
and proceeds to the following setup.
※
If One-point setup was chosen, the
setup is complete.
4
The upper-limit value
setup
The display now changes to
[H. SEt], indicating that the upper-limit
value can be set.
The currently set upper-limit value
flashes.
5
Loading the sample of the upper-limit value
Load the sample of the upper-limit value
on the measurement pan.
6
Saving the upper-limit value
Press the Memory key.
After saving the upper-limit value, the
balance displays it briefly and
terminates the setup.
Continuous pressing
Key released