WEIGHING
The scale does not switch on
•
Make sure the power cable is connected properly
•
Connect the battery charger and try again. If the instrument continues to malfunction, contact the dealer.
The scale does not carry out sampling
•
The weight is unstable (
Err.Mot
)
•
The weight is insufficient, add more pieces and try again (
Error
)
The scale switches off suddenly
•
Automatic switching off active
•
Low battery
•
Battery failure
•
Power supply line failure
The scale is not reactive
•
One of the available energy saving modes has been activated
•
An unsuitable weighing filter has been selected
The weight is unstable
•
Check the active weighing filter (see page 47).
•
If the support surface is subjected to vibrations from machinery or moving vehicles, move the scale onto another
surface and try again.
The scale display switches off and displays a dot
•
Stand-by mode is active: press a key to reactivate weighing.
•
Energy saving mode is active: contact the dealer for further details.
The scale displays a permanen “
Zero
” message
•
The scale is unable to automatically reset the weight because it exceeds the maximum resettable weight at
switch-on.
•
Unload the scale and try again. If the scale continues to have the same problem even when there is nothing
on it, contact the dealer.