CALIBRATION SETTINGS AND PROCEDURES
Name
Modbus
address
CANopen®
Index/sub-
index
Profibus cyclic
IN/OUT
Acyclic DPV1
slot/index
Type Access
Maximum capacity
0x000C
0x3002 / 0x00
(M)
R : 0x0420
W: 0x0421
0x02 / 0x00
Ulong RW
Number of calibration
segments
0x000E
0x3000 / 0x00
R : 0x0222
W: 0x0223
0x02 / 0x01
Uint
RW
Calibration load 1
0x000F
0x3001 / 0x01
(M)
R : 0x0424
W: 0x0425
0x02 / 0x02
Ulong RW
Calibration load 2
0x0011
0x3001 / 0x02
R : 0x0426
W: 0x0427
0x02 / 0x03
Ulong RW
Calibration load 3
0x0013
0x3001 / 0x03
R : 0x0428
W: 0x0429
0x02 / 0x04
Ulong RW
Sensor sensitivity
0x0015
0x3004 / 0x00
(M)
R : 0x042A
W: 0x042B
0x02 / 0x05
Ulong RW
Scale interval
0x0017
0x3003 / 0x00
R : 0x022C
W: 0x022D
0x02 / 0x06
Uint
RW
Zero calibration
0x0018
0x3005 / 0x00
R : 0x0434
W: 0x0435
0x03 / 0x04
Long RW
Span coefficient 1
0x001A
0x3006 / 0x04
R : 0x0436
W: 0x0437
0x03 / 0x05
Float RW
Span coefficient 2
0x001C
0x3006 / 0x05
R : 0x0438
W: 0x0439
0x03 / 0x06
Float RW
Span coefficient 3
0x001E
0x3006 / 0x06
R : 0x043A
W: 0x043B
0x03 / 0x07
Float RW
Span adjusting coefficient
0x0020
0x3006 / 0x01
(M)
R : 0x042E
W: 0x042F
0x03 / 0x00
Ulong RW
Calibration place g value
0x0022
0x3006 / 0x02
R : 0x0430
W: 0x0431
0x03 / 0x01
Ulong RW
Place of use g value
0x0024
0x3006 / 0x03
R : 0x0432
W: 0x0433
0x03 / 0x02
Ulong RW
Zero offset
0x0092
0x2500 / 0x00
(M)
R : 0x0472
W: 0x0473
0x03 / 0x03
Long RW
11.1
Principles
eNod4
is factory calibrated to deliver
500 000 counts for 2mV/V
with a load cell on the
A3
input connector.
The measurement scaling in
eNod4
can be adapted to his application by the user. Some settings and the 2 calibration
methods allow the user to define his specific span according to his sensors characteristics.
When using eNod4 for legal for trade purpose, it is
imperatively required to activate the legal for trade switch
BEFORE any calibration procedure (cf § legal for trade switch
)
.
!