SBS-2500 Manual 09-15-TI2500
Note
The unit display flashes when the difference in temperature between the sample and the ambient air is greater
than ±5 °C
For samples of high viscosity (> 2000 mPa
•
s):
•
Use an external syringe to fill the measuring cell.
6.2 Filling the Measuring Cell Using the Built-In Sample Pump
•
Ensure that the sample tube is in the sample.
•
Press the fill button (12) slowly to fill the measuring cell.
•
Ensure that there are no air bubbles in the measuring cell.
6.3 Filling the Measuring Cell Using an External Syringe
•
Press the drain button (3).
•
Slide up the lock button (11).
•
Remove the screw stopper (6) with a coin.
•
Screw in the adapter for the external syringe.
•
Hold the sample tube over a waste container.
•
Slowly inject the sample into the measuring cell.
•
Ensure that there are no air bubbles in the measuring cell.
•
Do not remove the syringe while measuring.