26
Port
Enclosure
Range freq. (Hz)
50
Level (A/m)
30
Reference Standard
EN 60601-1-2
Power frequency magnetic fields immunity test
Reference document:
Level:
Test Equipment:
Environmental Conditions:
Uncertainty:
EN 61000-4-8
EN 60601-1-2
Generator PMM DAS-G120, Coil DANA DAS S120
Temperature: 20°C, RH: 40%
<8%
SUMMARY OF TESTS RESULTS
The following table summarizes the tests performed:
Port
Enclosure
Enclosure
Enclosure
Enclosure
Enclosure
Frequency range
(MHz)
30-1000
.............
80-2700
See point 6.2.3
0.00005
Test Level
Class B
±15KV air
±8KV contact
3V/m modulated
AM 80% 1KHz
See point 6.2.3
30A/m
Test Type
Radiated emission
Electrostatic discharge
immunity
Radiated electromagnetic
field immunity test
Immunity to RF wireless
communication
equipment
Power frequency
magnetic fields
immunity test
Reference
Standards
EN 60601-1-2
EN 60601-1-2
EN 60601-1-2
EN 60601-1-2
EN 60601-1-2
Result
Ok
Ok
Ok
Ok
Ok