Materials : Sensor Body : Acetal Copolymer, TICONA M90
Turbine : Acetal Copolymer, TICONA M90
Stick of Turbine : #304 Stainless Steel
O-Ring : EPDM Rubber
Housing : ABS resin
5.0
Operation Guide
5.1 Install Batteries
Slide off the cover case of the battery room, and put into 2 AAA batteries. Then slide the cover case
back.
5.2 Connecting Fitting
The specification of the DigiFlow 6000R connectors is 3/8” MNPT female. You could use any 3/8”
MNPT male connectors.
5.3 Setting Filter Capacity
Keep pressing DISPLAY button, then press RESET button and release both of the buttons. The
DigiFlow 6000R will go into setting procedure. The digit will blink “FLT 1” four times, then
blink “ OFF” to notice you that the setting procedure is on. Press DISPLAY button to choose the
capacity value. The sequence is OFF - 0 – 100 – 200 – 300 …… 19800 – 19900 – OFF – 0 –100
(gallon model) and OFF - 0 –500 – 1000 --- 99500 – OFF – 0 – 500 (Liter model). Then press
RESET button to save the filter capacity value.
* If the setting value is “OFF”, the monitor function of the dedicated filter will be disable.
5.4 Operating
The DigiFlow 6000R will be automatically turn on when there is water flowing through the sensor.
It will automatically turn off in 10 seconds when there is no water flowing. The DigiFlow 6000R
display the total flowing volume.