TD_US-BA-e-2020
5
3. The Plus key 3-6 or the Minus key 3-9 is pressed to select the material code for
measuring. Then the material selection button 3-5 is pressed to confirm. The
measuring instrument changes to the measuring mode, '0' appears in the
display. If a material code is selected without confirming this selection, the
instrument will automatically return to the measuring mode after a few seconds.
In this case, the meter will still retain the old material code before switching off
completely.
4. A
4-digit
number
is
shown
on
the
display:
If the plus key 3-6 is pressed, it will display 'cd11' or if the minus key 3-9 is
pressed, it will display 'cd01'. The 4-digit number represents the last ultrasonic
speed
defined
by
the
user.
By changing the ultrasonic velocity, varying material properties can be
compensated.
5. Once the material code has been selected and stored, it is stored in the
instrument memory. As long as no change is made, the meter always uses this
material code.
6. To enter the material code selection menu, press the selection button 3-5. To
leave the menu again, press the selection button 3-5 again or wait until the
device - after a few seconds - returns to the measuring mode and the display
shows '0'.
5.
Calibration (Adjustment)
1. A little contact gel is applied to the zero plate 3-12.
2. The calibration button 3-8 is pressed and 'CAL' appears in the display. CAL' is
the abbreviation for calibration.
3. The sensor 3-7 is pressed onto the zero plate. The display "Measuring process
in progress"
((
•
))
is shown when the measuring process has been successfully
established by the transmit & receive process. The display shows '5.0' mm
(target thickness of the zero plate) and 'CAL' alternately. As soon as the display
value has stabilized, press the 'CAL' key 3-8 to confirm. Then the instrument
switches back to the measuring mode.
4. The calibration (adjustment) is now completed and is automatically stored in the
device.