Instruction Manual 

TB 

 

TB-BA-e-1212 

 

1

 

DIGITAL COATING THICKNESS GAUGE  

 

 

  

Models:  

TB 1000-0.1 F 

                 TB 1000-0.1N 
                 TB 1000-0.1FN 
                 TB 2000-0.1F 
            

Table of contents 
 

1. 

Features 

2. 

Specifications 

3. 

Front panel descriptions 

4. 

Measuring procedure 

5. 

OFFSET-

Accur function 

6. 

Calibration 

7. 

Battery replacement 

8. 

Calibration foils 

9. 

Correct handling of coating thickness 
measurement with external sensors 

       10.   Considerations/ Trouble-shooting 
       11.    Declaration of conformity 
 

Annotation: It is strongly recommended to calibrate  
                     the new instrument before the first use, as 
                     described in paragraph 6. By doing this it  
                     will be achieved a much better measure- 
                     ment result right from the start. 

 

1. Features 

                             
*    The instrument meets ISO standards 2178. So it can   

be used in laboratory conditions as well as 
under “harsh field” conditions.  

*    The instrument is widely used to measure the thickness 
      of these nonmagnetic coatings: paint, plastic & PVC, 
      porcelain enamel, copper, zinc, aluminium, chrome, 
      layers of laquer, layers of paper etc.   
      These coatings may be put up on following magnetic  
      base grounds: steel, iron, nickle, etc.  
*    Wide measuring range and high resolution 

 
2. Specifications 

                             

Display:

   

4 digits, 10 mm LCD 

 

Range:

      

1000 

μm and 2000µm 

Resolution:

 

0.1 

μm to 100 μm 

                                1 µm over 100 µm 
 

Accuracy: 

 
Following accuracy is given for  the models TB 1000-0.1F, 
TB 1000-0.1N as well as  TB 1000-0.1FN: 
 
- Standard:   3% of the measured value or min. ± 2.5 µm 
Is valid within a tolerance range of ± 100 

μm around the 

individually measured range, if a two-point calibration was 
performed within this tolerance range. 
 
For model TB 2000-0.1F accuracy is: 
 
- Standard:   5% of the measured value or min. ± 2.5 µm 
Is  valid within a tolerance range of ± 100 

μm around the 

individually measured range, if a two-point calibration was 
performed within this tolerance range. 
 

Off-Set Accur

 Mode:   1% of the measured value 

                                      or min. ± 1.0 µm    
  Is valid within ± 50 

μm around the 

Off-Set Accur

 point.  

 
All tolerances are valid after the adjustment procedure! 
                            
Power supply: 

4 x 1.5V AA battery 

 
Operating conditions: 

 

 

Temperature  

0 to 50 °C 

 

Humidity   

≤ 80 % 

 
Size:              

161 x 69 x 32 mm 

 
Weight:   

about 260 g (including batteries) 

 
Accessory: 

Carrying case  

 

 

Operation manual 

 

 

Instrument and sensor 

 

 

Calibration foils 

 

 

Base plate  

 

3. Front panel description 
 

3-1 Sensor 
3-2 Display 
3-3 Zero- key 
3-4 Plus- key 
3-5 Minus- key 
3-6 Power- on/ Power- off key 
3-7 µm/ mil key 
3-8 Battery compartment/ cover 

 