11.1.5 Switching tones on and off
Menu Unit
→
settings Switch tones
→
on and off
11.1.6 Service
Menu Unit Settings
→
Service
→
11.1.6.1 Display raw data
Switch here to display the raw measured values during adjustment as AD value or as
voltage in mV.
11.1.6.2 Reset unit
Do not use this function! Contact the service department!
11.1.7 Device info
Menu Device Settings
→
Device Info
→
Here they receive information about
- Device name or model
- Installed firmware revision
- Serial number of the unit
11.2 Measurement menu Setting
11.2.1 Measuring frequency
11.2.1.1 Setting
By clicking on change, you can change the frequency with which the measured values
are stored in the "Track" mode or transmitted to the interface! The following frequencies
are possible: 1000, 500, 100, 50, 10, 5 and 1 Hz.
11.2.1.2 What frequency is needed?
Choose the frequency as low as possible and as high as necessary! High frequency
data storage or transmission causes a large amount of data which takes a long time to
transmit.
11.2.1.2.1 For long-term measurements
such as a fatigue test, select a very low frequency, max. 50Hz.