Manual 

 

 

 

SAUTER FS 

V. 1.0 
01/2022

 

EN 

 

 
 

 

 

                                                    

 

FS-BA-en-2210.docx 

Measuring channel CH2 
Measuring channel CH1 
Interface socket  
(optional) 

Measuring channel  
CH4 (optional) 
 
Measuring channel 
CH3 (optional) 

Содержание FS 2

Страница 1: ...ohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www sauter eu Manual SAUTER FS V 1 0 01 2022 EN FS BA en 2210 docx Measuring channel CH2 Measuring channel CH1 Interface socket optional Measuring channel CH4 optional Measuring channel CH3 optional ...

Страница 2: ...lectrostatic sensitive device 5 4 3 Risk of accident 6 4 4 Damage defective equipment 6 4 5 Accuracy 6 4 6 Maximum measuring range 7 5 Available models 7 6 Scope of delivery 8 7 Connections 8 7 1 USB C socket 8 7 2 Connection sockets for external sensors 9 7 2 1 Compatibility and necessary accessories 9 7 2 2 Plug play from external sensors 9 7 2 3 Pin assignment connector 10 7 3 Expansion from 2 ...

Страница 3: ...te 17 11 1 3 Set energy saving options 17 11 1 4 Display brightness 17 11 1 5 Switching tones on and off 18 11 1 6 Service 18 11 1 7 Device info 18 11 2 Measurement menu Setting 18 11 2 1 Measuring frequency 18 11 2 2 Readability 19 11 2 3 Functions 20 11 2 4 Channel limits 21 11 3 Sensor settings 22 11 3 1 Adjustment 22 11 3 2 Calibration data 23 11 3 3 Overload values 23 11 4 Internal memory 24 ...

Страница 4: ...Power supply 27 14 1 Charging the battery 27 14 2 Plug in power supply unit 28 14 3 Power supply via PC 28 15 Interface 28 15 1 Interface description 28 15 1 1 USB 28 15 2 Interface protocol 28 16 Declaration of conformity 28 ...

Страница 5: ...r even the unit itself has been damaged If any damage is apparent please notify the specialist dealer or manufacturer immediately 3 Intended use The SAUTER FS device is intended for measuring forces and torques in industrial environments 4 Warnings 4 1 Transport A lithium ion battery is built into the unit The technical data of the battery is described in chapter 8 1 is described Observe the natio...

Страница 6: ...ads of the accessories you are using Check measuring cells regularly for deformations and cracks Shocks vibrations and torsion on the force application bolt can damage the sensor Only transport the unit in the transport case supplied 4 4 Damage defective equipment Do not use damaged measuring instruments cables plug in power supplies or similar Contact the service department of the dealer or manuf...

Страница 7: ...ed Overload values are stored Overloaded units and sensors are excluded from warranty return and exchange 5 Available models Model internal measuring cell Measuring range internal measuring cell Max N Readability internal measuring cell N Number of measuring channels Max possible resolution per channel d FS 2 2 10 000 FS 2 20 20 0 004 2 10 000 FS 2 50 50 0 01 2 10 000 FS 2 100 100 0 02 2 10 000 FS...

Страница 8: ...e EU plug in power supply unit Operating instructions 7 Connections 7 1 USB C socket Your unit is supplied with power via the USB C socket on the right hand side of the unit and serves as an interface for data transfer to a PC The interface description can be found in chapter 15 Interface chapter USB C ...

Страница 9: ...corresponding necessary services The services listed in chapter 13 2 include the adjustment of the measuring chain the connector and the required data memory chip 7 2 2 Plug play from external sensors External sensors can be unplugged and replaced by other adjusted sensors Due to the tolerances of the respective input channels the sensors must be adjusted with the device used For this purpose the ...

Страница 10: ... socket is available on the unit Interface currently not available 8 Technical data 8 1 Battery Lithium ion rechargeable battery Nominal voltage 3 7V Capacity 2400mAh or 8 88Wh Safety data sheets are available in the manufacturer s webshop or on request Do not use devices with defective batteries Do not send defective batteries or devices with defective batteries Pin Connection 1 N C 2 Memory chip...

Страница 11: ...l Measurement accuracy o with internal measuring cell 0 1 of Max o with external measuring cell among other things from the measuring cells used Overall dimensions W D H 71 31 180 mm Overload protection 150 of Max with internal measuring cell Thread on load receptor M6 external Internal battery operation standard operating time up to 8 h charging time approx 8 h External mains adapter for connecti...

Страница 12: ... hold the ON OFF button for approx 3 sec 9 2 Activate display with activated energy saving options If you have activated the energy saving mode of the display as described in section 11 1 3 it is possible that your unit is switched on and only the display is switched off To activate the display briefly press the START STOP key or touch the display If the display cannot be activated start the machi...

Страница 13: ...isplayed instead 9 4 Measuring with external sensors The external measuring channels are displayed as CH1 and CH2 for FS 2 x and CH1 to CH4 for FS 4 x On the external channels CH1 to CH4 it is possible to measure and save the data simultaneously 9 5 Plugging in and unplugging external sensors To plug in or unplug the external sensors the meter must be switched off or must be in the main menu press...

Страница 14: ...old the button for at least 0 3 seconds to set all connected sensors to 0 10 5 Save measurement data The different measuring functions that you can use and how the measured values are stored is described in section 11 2 3 10 6 Force direction Units with internal sensor can measure forces in tension direction positive measured value and in compression direction negative measured value External sens...

Страница 15: ...d transverse forces These cause measurement errors and can damage the sensor Ideal force application Incorrect force application 11 Main menu settings When your device is in the start menu you can make settings by clicking on Menu This will take you to the main menu with the following sub pages wrong wrong wrong ideal ideal ideal ...

Страница 16: ...revious page Press the ON OFF button to return to the start screen 11 1 Menu unit settings Device settings Measurement settings Sensor settings Internal memory Main menu Device Language Date Time Service Unit information Energy saving options Display Audio ...

Страница 17: ... OK 11 1 3 Set energy saving options Menu Machine Settings Energy Saving Options Set here when the unit or display should switch off to save energy Switching off the display is only recommended for long term measurements As a rule never switch off the unit and display To reactivate the display briefly press the START STOP button or touch the display 11 1 4 Display brightness Menu Device settings D...

Страница 18: ...el Installed firmware revision Serial number of the unit 11 2 Measurement menu Setting 11 2 1 Measuring frequency 11 2 1 1 Setting By clicking on change you can change the frequency with which the measured values are stored in the Track mode or transmitted to the interface The following frequencies are possible 1000 500 100 50 10 5 and 1 Hz 11 2 1 2 What frequency is needed Choose the frequency as...

Страница 19: ...𝑔 𝑑 𝑒𝑖𝑛𝑔𝑒𝑠𝑡𝑒𝑙𝑙𝑡𝑒 𝐴𝑏𝑙𝑒𝑠𝑏𝑎𝑟𝑘𝑒𝑖𝑡 𝑖𝑛 𝑁 𝑆𝑇𝐸𝑃 𝐴𝑏𝑙𝑒𝑠𝑏𝑎𝑟𝑘𝑒𝑖𝑡𝑠𝑓𝑎𝑘𝑡𝑜𝑟 𝑑𝑚𝑖𝑛 𝑀𝐴𝑋 𝑛 𝑑 𝑑𝑚𝑖𝑛 𝑆𝑇𝐸𝑃 𝑀𝐴𝑋 𝑛 𝑆𝑇𝐸𝑃 Example FS 2 100 𝑀𝐴𝑋 100 𝑁 𝑑𝑚𝑖𝑛 100𝑁 10 000 0 01𝑁 𝑆𝑇𝐸𝑃 2 𝑑 𝑀𝐴𝑋 𝑛 𝑆𝑇𝐸𝑃 100𝑁 10 000 2 0 02𝑁 Set the readability factor as high as possible and as low as necessary Very low readability often leads to unstable display values with the measuring technology used and the high sampling frequencies Readability fact...

Страница 20: ...art Stop button for more than 3 seconds to end the recording in the displayed file Recording finished display 6 To save a new measurement series in a new file proceed again as in point 1 11 2 3 2 Peak To detect a peak value select the Peak setting When the measured value is shown in the display start the detection by briefly pressing the START STOP button Please note that a peak can only be detect...

Страница 21: ...STOP button again the mode starts from the beginning and a new file is created 11 2 3 4 Manual with statistics The manual storage of the measured values is the same as described in chapter 11 2 3 3 described In addition the min max and average values of the stored measured values are displayed here 11 2 4 Channel limits This setting controls the traffic light displayed when you are working in the ...

Страница 22: ...d of the sensor to be adjusted 11 3 1 1 Measuring range and unit Do not change these values A new adjustment and possibly also a recalibration will be necessary 11 3 1 2 Readability The displayed value is the smallest adjustable readability Set the readability required for your application 11 2 2 chapter set the readability required for your application before using the meter 11 3 1 3 Two point ad...

Страница 23: ...n chapter 10 described 11 3 2 Calibration data On this page you will find information if a measuring device with internal cell CH0 or a device with external measuring cell CH1 to CH4 has been calibrated in the KERN laboratory The specified measuring channel indicates to which channel CH1 to CH4 an external sensor was connected during adjustment and calibration It is recommended to always use exter...

Страница 24: ...end the selected files to the PC The optionally available plug in for MS Excel SAUTER AFI 2 0 is recommended for this purpose 11 4 3 Delete file Select the files you no longer need to delete them Do not collect data on the device Transfer them to another storage medium as soon as possible and delete the entire data memory of the FS 11 4 4 Read out files Reading out files with a PC is not possible ...

Страница 25: ...E 04 Strap tension clamp Clamping width from 0 to 6 mm 0 to 500 N AE 05 Rope and thread tension clamp 0 to 500 N AE 06 Cable pull off clamp Clamping width 1 5 to 6 mm 0 to 500 N AE 07 Wedge clamp Clamping width 0 to 6 mm 0 to 500 N AE 2K screw clamp Clamping width 0 to 15 mm 0 to 2000 N AE 500 Screw clamp Clamping range from 0 to 10 mm 0 to 500 N AFK 02 Holder for force gauges Further accessories ...

Страница 26: ...METER MEMORY mounted on measuring cell and two point adjusted with weight up to 250kN 13 2 1 Adjust and calibrate In order to adjust and calibrate the entire measuring chain a measuring device and sensor are required in the core laboratory Your QM representative can provide you with information on the period in which recalibration is necessary In any case we recommend a DAkks calibration certifica...

Страница 27: ...red values with EXCEL plug in With the optionally available EXCEL plug in SAUTER AFI 2 0 stored measurement data can be transferred to a PC EXCEL software not included in delivery 14 Power supply 14 1 Charging the battery The device is equipped with a rechargeable battery at the factory Charge the battery using the supplied USB C cable and a power source with the following values Voltage 5V DC Cur...

Страница 28: ...a via the USB interface 15 Interface 15 1 Interface description 15 1 1 USB This is a virtual USB interface The device is not recognised as a storage medium Baud rate 115200 Databits 8 Stopbits 1 Parity none 15 2 Interface protocol The interface protocol KCP KERN Communication Protocol can be downloaded from the website www kern sohn com Important ASCII interface commands SI Sends the current readi...