![sauermann VT 210 Скачать руководство пользователя страница 22](http://html1.mh-extra.com/html/sauermann/vt-210/vt-210_user-manual_1202769022.webp)
8.4 Hygrometry probes
8.4.1 Unit
Temperature:
➢
Go to the line
“Temperature”
and press
OK
.
➢
Select the required unit: °C or °F
➢
Press OK to validate the unit selection.
8.4.2 Alarm
Temperature:
➢
Go to the line
“Temp. alarm”
and press
OK.
➢
Select
“High alarm”
and/or
“Low alarm”
by pressing
OK
.
It is possible to set the high and low thresholds.
➢
Go to the line
“High threshold”
and press
OK
.
➢
Set the threshold with the arrow keys between -9999.9 and 9999.9°C then press
OK
.
➢
Go to the line
“Low threshold”
and press
OK
.
➢
Set the threshold with the arrow keys between -9999.9 and 9999.9°C then press
OK
.
8.4.3 Atmospheric pressure
➢
Go to the line
“Atmo. Press.”
and press
OK
.
➢
Select the atmospheric pressure with the arrow keys between 800 and 1200 hPa.
➢
Press
OK
to validate.
8.5 Tachometry
8.5.1 Unit
➢
Go to the line “
Tachometry”
and press
OK
.
➢
Select the required unit: tr/min or RPM.
➢
Press OK to validate the unit selection.
8.5.2 Type
➢
Go to the line
“Tacho type”
and press OK
.
➢
Select the required type: Optic or Contact
➢
Press OK to validate the type.
8.6 Multifunction probe
8.6.1 Unit
Air velocity:
➢
Go to the line
“Air velocity”
and press
OK
.
➢
Select with the arrow keys the required unit: m/s, fpm, km/h or mph.
➢
Press
OK
to validate the unit selection.
Temperature:
➢
Go to the line
“Temperature”
and press
OK
.
➢
Select the required unit: °C or °F
➢
Press OK to validate the unit selection.
Airflow:
➢
Go to the line
“Airflow”
and press
OK
.
➢
Select with the arrow keys the required unit: m
3
/h, L/s, cfm or m
3
/s
22
SETTING OF MEASUREMENT PARAMETERS