sauermann Kimo TX Series Скачать руководство пользователя страница 3

3  Connection of the liquid column manometer

•  For a pressure measurement: connect the crystal tube on the left connection (+).
•  For a depression measurement: connect the crystal tube on the right connection (-).
•  For a differential pressure: Connect on the left connection (+) the highest pressure 

and on the right connection (-) the lowest pressure.

•  There is not enough liquid in the receiver, in this case : add some liquid (see step 1). 

Screw the pressure connection and wait 2 to 3 minutes, then repeat the procedure to 

set the zero as described hereabove.

•  There is too much liquid, in this case : take an empty burette, squeeze it and suck up 

the liquid excess. Screw the pressure connection and wait 2 to 3 minutes then repeat 

the procedure to set the zero as described hereabove.

Correct reading : 0.7 mmH

2

O

This handling can cause an important move of the ruler, two 

problems may occur:

Connect the manometer with a Ø 5 x 8 mm crystal tube to the source of 

pressure or depression to check:

If you need pass through a wall to make some pressure measurements, it is 

advised to use glanded penetrations in order to avoid a premature wear of 

the silicone tubes (there are different lengths available).

4  Calibration of a manometer

Example: 

If the generator shows 4.8 mmH

2

O and the manometer 5 mmH

2

O: in this case, slightly 

unscrew the setting screw of the level until getting the same value (the bubble of the 

spirit level must be between the two lines. See step 1). 

If the generator shows 5 mmH

2

O and the manometer 4.8 mmH

2

O: in this case slightly 

screw back on the setting screw of the level until getting the same value (the bubble of 

the spirit level must be between the two lines. See step 1). 

•  Redo the zero point of the manometer and wait for the stabilization of the 

liquid (2 to 3 minutes).

•  With a pressure generator, generate a determined 

pressure.

•  Wait 2 to 3 minutes the stabilization of the 

measurement.

5 Maintenance

The manometers do require no particular maintenance. Nevertheless, we recommend to 

change the manometer liquid once or twice a year.

•  Compare the results of the pressure generator and of the manometer. If the manometer 

has a different value from the value of the pressure generator, adjust the manometer 

with the setting screw of the level until getting the same value.