If there is a temperature gap between the electrode and the compost, wait for 30 seconds to remove the 
gap to prevent errors. Also, keep the same pressure to the sensor to stabilize the reading.

 

5.  Press [

測定

] (“Measure”, rightmost key). 

6.  For accurate reading, measure the same point repeatedly and press [

平均

] (“Average”, key above + sign) to show 

an average. 

7.  Clean the electrode after the measurement. 
8.  The device turns off automatically when one minute has passed. 
 
 

Setting of Customized Formulae 

 
Each angle and equilibrium of five conversion formulae can be modified individually. Formula A can also be edited, 
but it is advised to remain it. 

Angle change 
   
1.  Press [

選択

] (“Select”, key above − sign) to choose the formula 

to be edited. 

2.   Press [

変更

/

決定

] (“Set”, lower left key). 

 

A character of “A” appears at the upper right, and the current 
angle is shown. The default value is “100” in all formulae. 

3.  Press + or − key to modify the value, available between 0 and 

199. The value increases or decreases continuously by a long 
press for more than 1 second. 

 

In the right case, set it to 66 referring to the equation A*. 

4.  Press  [

変更

/

決定

] (“Set”, lower left key) to complete,  or go to 

the next step if you also change the equilibrium. 

 
Parallel change 
 
1.  Press  [

測 定

]  (“Measure”,  rightmost  key)  when  the  angle  is 

displayed.   

      The character at the upper right changes to “H” and the current 

equilibrium is shown. The default value is “0.0” in all formulae. 

3.  Press + or − key to modify the value, available between −19.9 

and 19.9. The value increases or decreases continuously by a 
long press for more than 1 second. 

 

In the right case, set it to −3.0. 

4.  Press [

変更

/

決定

] (“Set”, lower left key) to complete, or press 

[

測定

] (“Measure”, rightmost key) to go back to angle change. 

5.  If  a  measured  value  was  displayed  before  the  setting,  a 

recalculated value is displayed. 

 
 

Specifications 

 
Name:   

Moisture Meter for Compost Model SK-950A 

Method:   

Electric resistance measurement 

Measuring range:  23.0 to 80.0% for Formula A, “H” or “L” is displayed above/below the range 
 

 

23.0 to 99.9% for customized formulae, “H” or “L” is displayed above/below the range   

Temperature compensation range: 0 to 80°C without freezing 
Power:   

One 9V battery 

Dimensions: 

Main body: 61 (W) x 146 (H) x 23 (D) mm 

 

 

Electrode: Ø 10 mm x 800 mm long 

 

 

Cable: 800 mm long 

Weight:   

143 g including a battery 

 
 

 

3-4, Kanda-kajicho, Chiyoda-ku, Tokyo 101-0045 Japan 

https://www.sksato.co.jp/en/

 

  Difference of known values          20%−10% 
Difference of measured values      25%−10% 

A* =                                                          =                      = 66% 

Plots 
of 20% 

Plots 
of 10% 

Measured 
value 

Known moisture content 

Set to 

66%*

 

Function 
after setting 

Measured 
value 

Known moisture content 

Function 
before 
setting 

Scatter 

plots 

Set to 
−3%

 

Ver. 1. As of Oct. 2021 