SATO SK-940A Скачать руководство пользователя страница 2

 

How to use 

1.  Check that the sensing part is clean. When the part is dirty, wipe it with a dried cloth. 

2.  Press [

測定

] (measurement) button and confirm that [L] displays. 

3.  By pressing [

選択

] (select) button, move the cursor to the range (S, 1, 2, 3, 4 and 5) you want to measure. 

4.  Apply the electrodes of probe to the sample (object to be examined).    Be sure to apply two electrodes to the 

sample with uniform force (pressure).           

5.  Press the [

測定

] (measurement) button either of the probe or main body.   

6.  For  accurate  readings,  measure  the  same  sample  more  than  three  times  and  detect  an  average  value  by 

pressing [

平均

] (average). 

7.  After the measurement, be sure to clean the sensor section. 

8.  Power will automatically turn off after 2 min.   

 
Average: 

By  pressing  the  [

平均

“Averaging” button,  averaged  moisture value of the up to last 9 data since the power has 

turned on is displayed.    When the 

“Averaging button” is pressed, the data stored are reset. If you measured more 

than 9 times, the data of the earlier measurement are erased and the latest 9 data are stored.    When the power is 
tuned off, all data stored are reset.   
 

 

NB: Be sure to press the averaging button before starting measurement of different sample. 
 

 
Notes on measurement

 

1.  It  is  desirable  that  there  are  little  difference  in  temperature  between  the  sample  and  the  unit  for  accurate 

measurement.     

2.  The  power  is  automatically  turned  off  if  the  unit  is  not  operated  for  approx.  2  min.    At  this  time,  all  stored 

data are reset. 

 

 
Display 

1.  [

平均

] (Average) 

 

“AVE” blinks at upper part of the display when measurement starts.   

      The latest 2 to 9 measurements are averaged.   

2.  Low battery mark   

 

“LOBAT” appears when the battery power is insufficient.   

      Replace the battery with new one. 

3.  Low moisture 

 

L” appears when the moisture content of sample is 10% or less,   

4.  High moisture content 

 

“H” appears when the moisture content of sample is 40% or more. 

 

 
Uses-range

 

(set up a moisture value to your original) 

It is possible to individually set the suitable value responds to each sample in 5 users-ranges. In the Users-ranges, 
range angle and equilibrium of range are changeable. 

A) Angle change   

1.  Select the range by pressing [

選択

] (select) button. 

2. Press [

変更/決定

] (Cal.) button.   

[A] appears at upper right and [100] is displayed. The default setting is [100] in 
all  ranges.  When  the  setting  was  changed  before,  the  value  of  the  latest 
setting is displayed.   

                        To lower the value, press [-] button 
                        To increase the value, press [+] button   

    NB: The movable range is 0 to 199.    Keep pressing the button 1 sec. 
            or more to change the numbers in succession. 
 
3.   Press [

変更/決定

] (CAL.) button again to complete the setup. 

S

1
2
3
4

LOBAT 

2. 

A) Angle change 

S

1
2
3
4

1. 

AVE9 

S

1
2
3
4

3. 

 

S

1
2
3
4

4. 

 

Отзывы: