INSTRUCTION MANUAL
No. 8312-00 Heat Stress Monitor
Model SK-170GT (JIS Conformed)
Страница 1: ...INSTRUCTION MANUAL No 8312 00 Heat Stress Monitor Model SK 170GT JIS Conformed...
Страница 2: ...idelines for heat stress Heat stress is caused not only by high temperature ambient but also by one s health condition one s adaptability to hot environment etc WARNING Never use this instrument in an...
Страница 3: ...uideline for proper working environment Features JIS B7922 class 2 compliant Compliant with JIS B7922 class 2 electronic Wet Bulb Globe Temperature WBGT index meter It can be used as a guideline for p...
Страница 4: ...the WBGT index reaches the limit set in advance Auto power off function Shuts down automatically in 20 minutes to save the battery Transmittable with RS232C Optional RS232C cable is available to send...
Страница 5: ...mode press the key for 2 seconds to set the wind speed After changing them press the key again for 2 seconds to fix the value and return to measurement mode IN_OUT Next Press this key for 2 seconds in...
Страница 6: ...TA globe temperature TG or humidity RH Also it displays the set value in alarm or wind speed setting mode MIN Indication for minimum measured value MAX Indication for maximum measured value WBGT Indi...
Страница 7: ...ction by turning the cover Turn on the power by pressing the POWER key Select a measurement mode by pressing the MODE Up key The display changes in the following order WBGT index WBGT Air temperature...
Страница 8: ...d speed to calculation For a better measurement set the ambient wind speed properly Note This instrument does not measure wind speed Press the POWER key to turn on Select the globe temperature TG mode...
Страница 9: ...out 5 8 8 0 m s 13 18 mph Wind moves small branches Wind raises dust and loose paper from the ground and drives them along Abstracted from the reports from National Weather Service U S Dept of Commerc...
Страница 10: ...r 20 minutes In case of using it for more than 20 minutes release the auto power off function Refer to Auto Power Off Function for details Note The buzzer does not work when MIN MAX values are display...
Страница 11: ...oftware exclusively made for this product Provide software by yourself when needed How to Connect Plug the RS232C cable into the connection jack on the right side of the unit Connect the 9 pin connect...
Страница 12: ...s missing Attach the globe to the main unit properly One of the measured values other than WBGT is over or under the display range Use within the measuring range E33 is displayed in the other measurem...
Страница 13: ...e ambient before use Electric noises generated from the environment affects the state Do not use the unit in the circumstances where electric noises are generated A value around 0 0 C blinks in TG mod...
Страница 14: ...pacitance sensor WBGT Alarm Setting Range 20 0 to 49 9 C 0 1 C step no deactivation Wind Speed Setting Range 0 1 to 3 0 m s 0 1 m s step Conformed Standard JIS B7922 Class 2 Approved measuring range w...
Страница 15: ...ght prepaid basis and correction of the defect will be made without charges We alone will determine if the product problem is due to deviations or customer misuse Out of warranty products will be repa...
Страница 16: ...3 4 Kanda kajicho Chiyoda ku Tokyo 101 0045 Japan https www sksato co jp en...