Errors Displays and Safety Instructions
If there is a malfunction detected, one of the following error codes will be displayed.
Code
Description
Solution
E02
The measured value is under the
display range.
Use within the measuring range.
E03
The measured value is over the
display range.
Use within the measuring range.
E04
The globe temperature sensor or the
internal electrical circuit could be
malfunctioning.
Stop using the instrument and contact the
dealer of your purchase or our company
for further assistance.
E11
The humidity sensor or the internal
electrical circuit could be
malfunctioning.
Stop using the instrument and contact the
dealer of your purchase or our company
for further assistance.
E33
The internal electrical circuit is
malfunctioning.
Stop using the instrument and contact the
dealer of your purchase or our company
for further assistance.
Reference Data
The Wet Bulb Globe Temperature (WBGT) is a composite temperature used to estimate the
effect of temperature, humidity and solar radiation (globe temperature) on humans. It is used
by industrial hygienist, athletes and the military to determine appropriate exposure levels to
high temperatures. It is derived from the following formulas.
Outdoors
WBGT= 0.7 x WB + 0.2 x TG + 0.1 x DB
Indoors
WBGT= 0.7 x WB + 0.3 x TG
WB = Wet bulb temperature
TG = Globe temperature
DB = Dry bulb temperature