Temperature Correction
An auto temperature correction has been attached in the models SK-100R, SK-101R, SK-102R,
SK-104R, SK-109R and SK-200R; however SK-106R and SK-107R do not have the function.
The SK-106R and SK-107R are also designed to measure samples when their temperatures are
20°C. For samples having other temperatures, perform temperature correction according to the
table below.
For example: for a sample of 25.0% at 24°C, 0.30% should be added. Therefore, the result
will be 25.3% (25.0 + 0.3 = 25.3 (%))
Temperature Correction Table (JIS B7185)
NOTES
- After measurement, clean the prism and the cover plate to completely remove the
sample, otherwise it may corrode the instrument.
- If the surface of the prism or cover plate is smeared with oil or similar liquids, it will repel the
sample and prevent measurement. Completely wipe off the contaminant with a soft material
such as tissue paper soaked in fresh water.
- Do not wash this instrument under running water.
- This instrument is an optical device.
Handle it with care, and avoid dropping or knocking it.