Figure 3.2 

– Left – post assembly for gradient measurement. 

                                   Right 

– post assembly for magnetic variation measurement. 

3.2. 

Processing Unit 

The processing unit allows for communication with an operator via a keyboard 

and a graphic LCD display. It contains circuits for processing and analyzing the 
measured signal: tuning circuits, low-noise amplifier, filter, and shaping amplifier. 

The processing unit is placed in a duralumin box. Two watertight bayonet 

connectors for connecting the sensors are located on the left side of the box. In AUTO 
mode, 

only the upper left connector is used, ‘Up sensor’. In MANUAL and WALK modes, 

the booth connectors can be used based on sensor configuration. The gradient is 
determined by subtraction: The value of the field measured in the location of the sensor 
connected to the ‘Up sensor’ connector minus the value of the field measured in the 
location of the sensor connected to 

the ‘Down sensor’ connector. 

There are two connectors on the right side of the box. The upper one serves to 

connect to the computer through USB. The connector for charging of the internal battery 
is located in the lower part of the instrument. The connector can also be used for an 
external power supply after removing the internal battery. 

The lead-acid battery is placed in the bottom part of the unit and is accessible by 

unscrewing the four bolts in the bottom cover. The battery compartment is hermetically 
separated from the electronics. 

The display and the keyboard are located on the top side of the unit. The 

keyboard contains 16 watertight membrane push-buttons. 

The instrument is delivered with a harness to fasten it on the chest of an operator. 

Содержание PMG-2

Страница 1: ...2021 SatisGeo s r o 1 29 2021 PMG 2 User Manual...

Страница 2: ...12 5 2 Tuning Of The Magnetometer 12 5 3 Setting The Backlight 14 5 4 Selection Of The Sensor Configuration 14 5 5 Walk Sample Rate Configuration 14 5 6 Setting The Time Interval To Switch Off 14 5 7...

Страница 3: ...2 8 1 USB Interface 19 8 2 Charger 19 8 3 Maintenance And Repair Work 19 8 4 Storage And Transportation 19 8 5 Warranty 19 8 6 Accessories 20...

Страница 4: ...ield WALK mode continually measures the magnetic field based on sensor configuration in three intervals 0 5s 1s 1 5s 2s 2 5s and 3s The possible sensor configuration is the same as in manual mode The...

Страница 5: ...ght 3 2 kg including battery Sensor Dimensions dia 80 x 200 mm Weight 0 7 kg Operating temperature range 10 to 60 C Storage temperature range 20 to 70 C The maximum gradient 1000nT m means that if a c...

Страница 6: ...l from the sensor and thus the size of the error of the measurement Figure 3 1 Post assembly for line measurement For measurements the sensor is mounted on a post which is assembled using up to five h...

Страница 7: ...ue of the field measured in the location of the sensor connected to the Up sensor connector minus the value of the field measured in the location of the sensor connected to the Down sensor connector T...

Страница 8: ...proportional to the measured magnetic field The entire measuring cycle is controlled by a microprocessor which computes the measured magnetic field and statistical error of the measurement from the va...

Страница 9: ...parameters 1 MEASURE item sub menu serves to activates measurement The possible measurements are MANUAL WALK and AUTO See chapter 6 2 POSITION Item serves for the setting of a number of a profile line...

Страница 10: ...ated in the setup menu item 5 S 08 04 28 DEVICE SETUP 1 BACKLIGHT 2 DATE TIME 3 OFF TIME 4 INFO 1 BACKLIGHT Item serves to switch the backlight on off or to set the time to elapse See chapter 5 3 2 DA...

Страница 11: ...10...

Страница 12: ...ivities of the operator on the keyboard or through the USB interface The interval can be set in the Setup menu item Time If the instrument does not display the above mentioned information even after r...

Страница 13: ...d magnetic field To display the current tuning interval select the TUNE sub menu MANUAL SET item in the setup menu For the SINGLE mode the display shows for example S 08 04 28 SET TUNE Sensor UP 41600...

Страница 14: ...00 43 800 nT 35 95 300 99 999 nT 18 43 500 45 800 nT Table 5 1 Tuning intervals A preliminary approximate setting of the tuning point can be performed on the basis of a magnetic field map see Figure 5...

Страница 15: ...ttery the instrument is automatically switched off after elapsing the set time interval after the last activities of the operator on the keyboard Thus the deep discharge and the damage of the battery...

Страница 16: ...and is changed by pressing 0 when the cursor is placed at the sign NEXT is used for switching among the items Line Pos and Step The modified data are stored in the memory by pressing ENTER Pressing ES...

Страница 17: ...sensor is changed and the statistical error of the measurement increases To achieve the maximum signal the sensor must be oriented in the north south direction so the north mark on the sensor must po...

Страница 18: ...displayed By pressing ESC the measured data is erased and the current position is displayed 6 5 Measuring In Walk Mode The WALK measure mode is selected in the MEASURE menu In this mode the instrument...

Страница 19: ...mation To store the data the operator presses ENTER otherwise ESC canceling the measured value and the measurement on the point can be repeated The position of the next measurement is displayed in bot...

Страница 20: ...netometer requires no special maintenance Only the built in lead acid battery needs care If the voltage drops under a certain level the battery might be damaged or the magnetometer might not operate c...

Страница 21: ...cessing Unit Console 1 pcs Sensor 2 pcs Harness 1 set Foot Stick 1 pcs Middle Stick 4 pcs Anchor Ring with Ropes 1 set Tent Pin 3 pcs Interface Cable to PC 1 pcs Charger 1 pcs Instruction Manual 1 pcs...

Отзывы: