22
EVO1S1 | SPO1S1 Operating Instructions
Operation
6.5.2 Weighing with Two Decimal Places (Not for Verified Models)
A menu setting is required for weighing with two decimal places (see
Chapter “6.7.2.4 Activating the Toggle Key (Not for Verified Models),” page 26).
t
Place the empty paint can on the weighing pan.
t
Briefly press the
0
key to zero.
y
The display shows “0.0 g.”
t
Press the toggle key
Z
.
y
The display shows “0.00 g.”
t
Add the first component “205.50 g.”
t
Read off the weight when the stability symbol (in this example) “g” is displayed.
t
Add the other components until the desired weight (formula) is reached.
t
Remove the filled paint can from the weighing pan.
When the scale is tared and the second decimal place with a resolution of 0.05 g is
activated via the toggle key
Z
, weighing up to 999.95 g with two decimal places
can be performed.
Values over 999.95 g can only be weighed using one decimal place.
Danger of Scale Damage!
Never close a paint can using a hammer while it is still on the weighing pan.
t
When closing the paint can, place it on a firm, stable surface.
6.6 Applications
Note for Scales for Use in Legal Metrology:
All application programs can be selected on scales verified for use in legal metrology.
Calculated values can be marked using the following characters:
− Percent = %
− Computed values = o,
l
symbol
0.0
0.00
205.50
213.50
593.85
1414.2