3
Getting Started
Remove the scale, load plate
and AC adapter from the
package.
Immediately after unpacking the
equipment, check it for any
visible damage.
Equipment Supplied:
Scale, load plate,
AC adapter, data cable
(25-pin - 9-pin)
Do not miss out on the benefits
of our full warranty. Please
contact your local Sartorius
office or dealer for further
information. If available,
complete the warranty
registration card, indicating the
date of installation, and return
the card to your Sartorius office
or dealer.
Choose a suitable place to set up the scale. Avoid exposure to
drafts, heat, moisture and vibration.
G
Check that the voltage rating of the AC adapter matches that
of your local line voltage (mains supply). If it does not match your
local line voltage, contact your Sartorius office or dealer. Use
only original Sartorius AC adapters! If you operate the scale in a
hazardous area/location outside Germany, you must comply with
the national electrical code and applicable safety regulations of
your country. For information on the legal regulations currently
applicable in your country, please ask your Sartorius office or
dealer or your paint supplier.
G
The model 609308... AC adapter must be installed outside
the hazardous area/location.
G
Before you operate your scale in a hazardous area/location,
the scale must be inspected either by a certified electrician or
under the guidance and supervision of a certified electrician to
make sure that the scale complies with the applicable regulations
(in Germany, in accordance with Section 12 of the ElexV).
Determine whether your scale must be reported to the technical
inspection authorities (e.g., trade board) in your country. The
system must also be inspected during operation. The system should
be inspected at intervals which allow for early detection of the
faults which occur as a result of normal wear and tear, so that they
can be corrected before damage is caused. In any case,
inspection must be performed at least every three years.
G
Fasten the AC adapter cable to the balance. Tighten down the
locking ring to secure the connection. Do not disconnect an
energized cable during operation in a hazardous location/area.
G
All applicable requirements (e.g., occupational safety
regulations) must be observed during operation.
G
Any tampering with the equipment by anyone, other than repair
work done by authorized Sartorius service technicians, will
invalidate the approval for use in hazardous areas/locations and