12 Sartolab
®
RF | BT Instructions for Use
Use
Requirements
−
The support manifold is suitable for the product (see Chapter 8.7.2,
page 20).
−
The product is equipped with a collection tube or a receiver bottle or is
connected to a receiver bottle.
Procedure
t
Insert the product adapter into the bracket of the support manifold
(see instructions for the support manifold).
4.5.2 Connecting the Product to the Vacuum Source via the
Vacuum Tube
Requirements
The product is
not
inserted in a support manifold.
Procedure
t
Attach the tube adapter (1) to the vacuum connection (2).
t
Attach the vacuum tube of the vacuum source to the tube adapter.
The vacuum tube
must not
come loose during filtration.
5 Use
5.1
Filling the Funnel
CAUTION
Danger of injury due to contamination of the vacuum source!
Excess liquid in the funnel can cause the liquid to be aspirated into the
vacuum source or the support manifold:
t
Do not
exceed the maximum filling volume (see Chapter 8.4.4,
Procedure
t
Make sure the funnel is securely attached to the collection tube or the
receiver bottle.
t
If required: Tighten the funnel.
t
Remove the lid from the funnel.
t
Pour the required quantity of liquid into the funnel.
t
If diatomaceous earth is used as a filter aid for clarifying filtration:
Observe the handling specifications (see instructions for the
diatomaceous earth filter aid).
t
Place the lid on the funnel.
1
2