
If you have selected “Automatic Output of Totalizing Application Data” in the
scale operating menu, you can document both the individual weights totalized and
the total of all weights when weighing individual samples.
Data output is generated automatically each time you press the
S
/
a
key.
Data Output
Transaction counter (1, 2, 3 etc.)
n
1
Net value
N
+
21.6 g
Automatic Output of Totalizing Application Data
Code
Changes
Accessible
3 13
1
Blocked
3 13
2 *
If the “Automatic Output of Totalizing Application Data” function is off, data output
is only generated when you press the
s
/
r
key (the current total is output).
Data Output
Transaction counter (1, 2, 3 etc.)
n
3
Current total net value
Sum-N +
62.8 g
With the totalizing function, you can totalize individual weight values (net, tare,
gross) or counting results that far exceed the actual capacity of the scale.
A partial amount is on the scale, but the total value is displayed.
Each time you store a value in the totalizing memory, the current transaction count
is displayed briefly.
When you press and hold the
S
/
a
key, “SUM” appears in the user guide
display and the current total (totalizing memory plus current weight value) is shown
in the main display.
* = factory setting
3–16
Содержание QC150FEG-S
Страница 37: ...1 32 Declaration of Type Conformity ...
Страница 38: ...1 33 EC Type Approval Certificate ...
Страница 125: ...4 22 V24 PC Scale Female connector PC 25 pin 9 pin 2 3 20 4 3 2 5 8 Buchse 25 pol oder 9 pol C H B D E ...
Страница 128: ...5 3 ...