Troubleshooting Guide
Error messages are displayed for approximately 2 seconds, after which the program automati-
cally returns to the weighing mode.
Problem
Cause
Solution
No segments appear
– No AC power available
– Check the AC power supply
on the weight display
– AC adapter is unplugged
– Connect AC adapter to power
supply
– Batteries are drained
– Replace batteries or recharge
them using an external battery
charger
oL
– The load on the pan
– Unload the scale
exceeds the scale’s capacity
UL
– Weighing pan is not positioned
– Correctly position the
weighing pan
– Contact between weighing
– Weighing pan may not come
pan and other objects
into contact with other objects
disErr
– Display overrun:
– Reduce load on the pan
Value to be output
cannot be displayed
calErr
– Calibration prerequisites
– Unload scale
were not met, e.g.,:
Zero/tare with the
t
key
– Zero/tare
Do not calibrate until “Zero" is
– Weighing pan is loaded
displayed
appErr
– Too little weight or, in the case – Increase weight
of application programs,
no paint can on weighing pan
PrtErr
– Interface port
– Contact Sartorius Customer
disabled for print function
Service responsible for
application programs
bAL.Err
– Scale is loaded
– Unload scale before
or defective when
turning it on
turned on
– Contact Sartorius Customer
Service
SYS.Err
– Scale is defective
– Contact Sartorius Customer
Service
Max. weighing
– Scale was turned on
– Position weighing pan
capacity less than
before positioning
– Turn the scale off and on
defined in chapter
weighing pan
again using the
e
key
on “Specifications"
The weight readout
– Scale is not calibrated
– Calibrate scale
is obviously wrong
– Scale was not tared
– Zero/tare scale
prior to weighing
If any other errors occur, please contact your local Sartorius customer service center.