19
Care and Maintenance
Cleaning
!
Unplug the AC adapter from the wall out-
let (mains supply). If you have a data
cable connected to the interface, unplug
it from the scale.
!
Do not use any aggressive cleaning
agents (solvents or similar agents),
concentrated acids or pure alcohol.
$
Make sure that no liquid enters the scale
housing
$
Clean the scale using either a paint brush
or a dry, soft and lint-free cloth.
Storage and Shipping Conditions
$
To ensure safe shipment, your scale has
been packaged using environmentally-
friendly materials.
You should retain these materials in case
you need to package your scale for
storage or return shipment.
$
Storage temperature: –20°C to +75°C
$
Permissible moisture level for storage of
the packaged scale: 90% max.
$
Read and follow the instructions given in
the section entitled “Safety Inspection.“
Safety Inspection
If there is any indication that safe opera-
tion of the scale with the AC adapter or
power supply is no longer warranted:
– Disconnect from AC power: unplug the
AC adapter/power supply from the wall
outlet (mains supply)
> Lock the equipment in a secure place to
ensure that it cannot be used for the time
being Safe operation of the PMA is no
longer ensured when:
– There is visible damage to the AC
adapter/power supply
– The AC adapter/power supply no longer
functions properly
– The AC adapter/power supply has been
stored for a relatively long period under
unfavorable conditions
– The equipment has been exposed to
rough handling during shipment
§
Observe the Warning and Safety
Information
In this case, notify your nearest Sartorius
Service Center or the International Tech-
nical Support Unit based in Goettingen,
Germany. Maintenance and repair work
may only be performed by service techni-
cians who are authorized by Sartorius
and who
– Have access to the required service and
maintenance manuals, or
– Have attended the relevant service
training courses
!
The seals affixed to this equipment
indicate that only authorized service
technicians are allowed to open the
equipment and perform maintenance
work so that safe and trouble-free
operation of the equipment is ensured
and the warranty remains in effect.