31
Repairing the Weighing Platforms
lastz_drehm071206.xls
P2020003.JPG
– Tighten the load cells with the required torque.
Torque Values
– After replacing load cells, you need to check the overload stops and adjust them
if necessary (see page 27-28).
AUT23705.JPG
AUT24012.JPG
Load cell designation
Torque value
SAG type 011469/220kg C3
130Nm
SAG type 011231/550kg C3
130Nm
SAG type 011232/1100kg C3
130Nm
SAG type 011233/1760kg C3
130Nm
MP58T / 91kg C3
130Nm
MP58T / 227kg C3
130Nm
MP58T / 454kg C3
130Nm
MP58T / 1134kg C3
130Nm
MP58T / 2268kg C3
130Nm
SAG type 011241/7.5 kg C3
10Nm
SAG type 011242/15 kg C3
10Nm
SAG type 011243/30 kg C3
10Nm
SAG type 011244/50 kg C3
14Nm
SAG type 011245/100 kg C3
14Nm
Load cell designation
Torque value
SAG type 011246/50 kg C3
14Nm
SAG type 011247/100 kg C3
14Nm
SAG type 011248/200 kg C3
14Nm
SAG type 011252/150 kg C3
35Nm
SAG type 011253/250 kg C3
35Nm
SAG type 011290/500 kg C3
35Nm
SAG type 011293/7,5 kg C3
14Nm
SAG type 011294/15 kg C3
14Nm
SAG type 011295/30 kg C3
14Nm
SAG type 011296/50 kg C3
14Nm
SAG type 011297/100 kg C3
14Nm
SAG type 011298/200 kg C3
20Nm
TEDEA 1510 / 100 kg
35Nm
TEDEA 1510 / 250 kg
35Nm
TEDEA 1510 / 500 kg
35Nm