Specifications | Spezifikationen | Spécifications | Especificaciones | Specifiche 

Table 1

 Specifications for Minisart

®

 RC|SRP|NY|PES with 4|15|25 mm membrane filtration area 

d

Housing material

Polypropylene (PP)

Membranes

– RC = Regenerated Cellulose 
– NY = Polyamide 
– SRP: hydrophobic PTFE = Polytetrafluoroethylene 
– PES = Polyethersulfone

Application limits

Max.  recommended  operating  pressure  4.5  bar | 65  psi                                                                          

Housing burst pressure

>7  bar | 102  psi

Max. temperature

121°C, 30 min (autoclavable)

Sterilization 

Non-sterile Minisart

®

 can be autoclaved or sterilized  

by ethylene oxide sterilization (EO)

 Minisart

®

 type

RC  

0.2 µm

RC  

0.2 µm

RC  

0.45 µm

SRP  

0.2 µm

SRP  

0.45 µm

NY  

0.2 µm  

NY  

0.45 µm 

PES  

0.2 µm

– Non-sterile packs:   
50 (K), 200 (S),  
500 (Q), 1000 (R) 
– sterile packs:  
individually packaged,  
50 (ACK)

K | S |  
Q | R

ACK

K | S |  
Q | R

K | S |  
Q | ACK

K | S |  
Q | ACK

K | Q |  
R | ACK

K | Q |  
R | ACK

K | Q | 
ACK

Bubble point (

>

)

with 
water 
3.0 bar| 
44 psi

with 
water 
4.6 bar| 
67 psi

with 
water 
2.0 bar| 
29 psi

with  
ethanol 
1.4 bar| 
20 psi

with  
ethanol 
0.9 bar| 
13 psi

with 
water 
3.0 bar| 
44 psi

with 
water 
2.0 bar| 
29 psi

with 
water 
3.2 bar| 
46 psi

Flow rate, 4 mm

 

d

 = 0.07 cm

2

 filter area | Hold-up volume

1

:  5–10 µl (

t

 ml/min)

with water at 1 bar

0.5

1.5

1.5

with methanol  
at 1 bar

1.5

3.0

2.0

4.5

4

with air at 0.1 bar

2

2

30

60

2

Flow rate, 15 mm

 

d

 = 1.7 cm

2

 filter area | Hold-up volume

1

:  30–100 µl (

t

 ml/min)

with water at 1 bar

20

10

40

20

40

40

with methanol  
at 1 bar

55

25

105

55

150

40

110

4

with air at 0.1 bar

2

2

2

800

1600

2

2

2

Flow rate, 25 mm

 

d

 = 4.8 cm

2

 filter area | Hold-up volume

1

:  100–200 µl (

t

 ml/min)

with water at 1 bar

80

50

160

50

100

100

with methanol  
at 1 bar

160

90

325

160

260

70

200

4

with air at 0.1 bar

2

2

2

1800

3000

2

2

2

Water penetration point –

> 4.0  bar| 
58 psi

3

> 3.0  bar| 
44 psi

3

Sterile filtration  
capability

5

 acc. to BCT

no

yes

no

yes

no

yes

no

     yes

Main applications

universal hydrophilic filter  
with very low non-specific 
binding, compatible with  
many aggressive hydrophilic 
solutions

universal hydro- 
phobic filter with 
very low non- 
specific binding  
suitable for almost 
all aggressive  
solutions, air and 
gases

hydrophilic filter 
with low non- 
specific binding, 
higher compati- 
bility with bases 
than RC

hydrophilic 
filter 
mainly for 
protein 
solutions, 
limited 
chemical 
compati-
bility

1

  Hold-up volume after air purge. For minimum hold-up please see “Note|Hinweis|Remarques|Nota|Note 1”

2

  Hydrophilic membranes can filter dry air or gas but become impermeable to air or gas when wetted!

3

  Hydrophobic membranes cannot be wetted with aqueous solutions unless you overcome their water  

penetration point or pre-wet them using an organic solvent (e.g. ethanol)

4

  PES is suitable for solutions only containing up to 30% MeOH

5

  According to bacterial challenge test (BCT) with 10

7

 

Brevundimonas diminuta. Non-sterile RC Minisart

®

 types 

are optimized for sample preparation and are not suitable for sterile filtration according to the BCT. All other 
non-sterile Minisart

®

 types listed above can be sterilized by autoclaving or EO before use for sterile filtration.

Содержание Minisart RC

Страница 1: ...Bei Spritzen mit einem Volumen kleiner als 10 ml wird schnell die maximal empfohlene Druck belastbarkeit von 6 bar berschritten Spritzenkolben von Spritzen kleiner 10 ml Volumen langsam herunterdr ck...

Страница 2: ...en using syringes with a volume of less than 10 ml as they can easily generate a pressure greater than the maximum recommended pressure resistance of 6 bar 87 psi for Minisart Therefore slowly press i...

Страница 3: ...pulse le reste de liquide qui se trouve l entr e et la sortie du filtre Cela permet de r duire le volume mort Attention 1 Avec des seringues d un volume inf rieur 10 ml on peut rapidement d passer la...

Страница 4: ...ida del filtro Esto reducir el volumen muerto Precauci n 1 Hay que tener cuidado al utilizar jeringas con un volumen inferior a 10 ml ya que se puede generar f cilmente una presi n mayor que la m xima...

Страница 5: ...e in uscita del filtro in questo modo si riduce il volume morto Attenzione 1 Con siringhe di volume inferiore a 10 ml la pressione operativa massima consigliata di 6 bar pu essere facilmente superata...

Страница 6: ...0 4 with air at 0 1 bar 2 2 2 800 1600 2 2 2 Flow rate 25 mm d 4 8 cm2 filter area Hold up volume1 100 200 l t ml min with water at 1 bar 80 50 160 50 100 100 with methanol at 1 bar 160 90 325 160 260...

Страница 7: ...45 m 200 17576 S 25 mm 0 45 m 500 17576 Q 15 mm 0 2 m 50 17558 K 15 mm 0 2 m 500 17558 Q 15 mm 0 2 m 50 17573 ACK 15 mm 0 2 m 50 17573 K 15 mm 0 2 m 500 17573 Q 15 mm 0 45 m 50 17559 K 15 mm 0 45 m 5...

Страница 8: ...ius Stedim Biotech GmbH August Spindler Strasse 11 37079 Goettingen Germany www sartorius com Nicht zur Wieder verwendung Single Use Ne pas r utiliser No reutilizable Monouso Bestellnummer Order Numbe...

Отзывы: