Levelling the Balance
Purpose:
– To compensate for unevenness at the place of
installation
Always level the balance again any time after it has
been moved to a different location. Only the 2 front
feet are used for leveling.
t
Screw in both rear support feet (only on models with
a rectangular weighing pan).
t
Turn the front levelingfeet as shown in the
illustration until the air bubble is centered within
the circle of the level indicator.
> Normally, several leveling steps are required.
t
On models with a rectangular weighing pan:
Screw out both back leveling feet until they touch
the setup surface.
Below-Cell Weighing
A port for a below-cell weighing hook is located on
the bottom of the balance.
y
Not permitted for applications in legal metrology.
t
Lift cover plate out of the bottom of the balance.
Attention: Place the balance on its side, do not turn
over completely!
t
Secure hook
1
: Use a wire, for example, to suspend
the sample on the hook.
y
Install a draft shield if necessary.
9
R
L
L
R
L
R
1