Error Codes
16
Error codes are shown on the main display for approx. 2 seconds.
The program then returns automatically to the previous mode.
Display/Problem
Cause
Solution
No segments appear
No power available
Check the power supply
on the display
The AC adapter is not plugged in
Plug in the AC adapter
Battery is drained
Replace battery; recharge
battery using external charger
oL
The load exceeds the
Unload the balance/scale
balance/scale capacity
UL
Weighing pan not in place
Place the weighing pan
on the balance/scale
Something is touching
Move the object that is
the weighing pan
touching the weighing pan
disErr
Display overflow:
Reduce load on the
Value cannot be shown
balance/scale
on the display
calErr
Calibration parameter not met;
Calibrate only when zero
e.g.:
is displayed
– Balance/scale not zeroed
Press
w
to tare the
balance/scale
– Balance/scale is loaded
Unload the balance/scale
appErr
Weight is too light or there is
Increase the weight
no sample on the balance/scale
on the balance/scale
with application in use
PrtErr
Data interface for
Contact the Sartorius
printing is blocked
customer service center
Bal.Err
Balance/scale loaded or defective
Unload balance/scale before
when power was turned on
switching on or contact
Sartorius customer service
Sys.Err
Balance/scale defective
Contact Sartorius customer
service
Max. weighing capacity The balance/scale was
Place the weighing pan
is less than indicated
switched on without
on the balance/scale and
under “Specifications”
the weighing pan in place
press
e
The weight readout
The balance/scale was not
Calibrate/adjust the
is obviously wrong
calibrated/adjusted before weighing balance/scale
Balance/scale not zeroed
Zero the balance/scale
If any other errors occur, contact your local Sartorius customer service center.