![Sartorius ED124S-0CE Скачать руководство пользователя страница 54](http://html1.mh-extra.com/html/sartorius/ed124s-0ce/ed124s-0ce_operating-instructions-manual_1199726054.webp)
Verified Models with EC-type Approval: Specifications
Model
ED224S-0CE,
ED124S-0CE
GK1203-0CE
GK703-0CE
ED224S-PCE
Type
BD ED 100
BD ED 100
BD ED 100
BD ED 100
Accuracy class
1
)
k
k
k
k
Weighing capacity, Max
1
)
220 g
120 g
1200 ct
700 ct
Scale interval d
1
)
0.0001 g
0.0001 g
0.001 ct
0.001 ct
Tare range (subtractive)
<
100% of the maximum capacity
Verification scale interval e
1
)
0.001 g
0.001 g
0.01 ct
0.01 ct
Minimum capacity, Min
1
)
0.01 g
0.01 g
0.1 ct
0.1 ct
Response time (average)
s
2.5
2.5
1.5
1.5
Range of use according
0.01 g – 220 g
0.01 g – 120 g
0.1 ct –
0.1 ct – 700 ct
to CD
1
)
1200 ct
Allowable ambient
°C
+17 to +27 (+63°F to +80°F)
operating temperature
Net weight, approx.
kg
4.8
4.8
4.7
4.7
Weighing pan size
mm
90
d
90
d
90
d
90
d
Weighing pan size
mm
230
230
160
160
Dimensions (W
+
D
+
H)
mm
230
+
303
+
330
230
+
303
+
260
Model
ED623S-0CE ED523S-PCE ED423S-0CE ED323S-0CE ED153-0CE,
ED153-PCE
Type
BD ED 200
BD ED 200
BD ED 200
BD ED 200
BD ED 200
Accuracy class
1
)
K
K
K
K
K
Weighing capacity, Max
1
) g
620
520
420
320
150
Scale interval d
1
)
g
0.001 0.001
0.001
0.001
0.001
Tare range (subtractive)
<
100% of the maximum capacity
Verification scale interval e
1
) g
0.01
0.01
0.01
0.01
0.01
Minimum capacity, Min
1
) g
0.02
0.02
0.02
0.02
0.02
Response time (average)
s
1
1
1
1
1
Range of use according
to CD
1
)
g
0.02 – 620
0.02 – 520
0.02 – 420
0.02 – 320
0.02 – 150
Allowable ambient
°C
+10 to +30 (+50°F to +86°F)
operating temperature
Net weight, approx.
kg
3.6
Weighing pan size
mm
115
d
Dimensions (W
+
D
+
H)
mm
230
+
303
+
136
1
) CD = Council Directive 90/384/EEC for non-automatic weighing instruments;
applicable to the European Economic Area
54