10
Instruction Manual Centrifuge SMC 6 V
7.3. Loading the rotor
The centrifuge must be evenly loaded for it to function properly.
To ensure even loading, comply with the following rules when loading the
swing-out rotor for six tube holders:
• Opposite-located tube holders must be identical.
• Opposite-located tube holders must either be empty or loaded with samples of
the same weight.
• When centrifuging an uneven number of samples, insert a tube filled with water
as a counterbalance.
7.4. Centrifugation
If a problem occurs during operation which makes it necessary to switch off the centrifuge,
push the “OPEN/STOP” button immediately.
Procedure:
• Press “OPEN/STOP” button to unlock the lid.
• Insert sample tubes in the tube holders. Observe rules for even loading! The most recent
program selected lights up.
• Press “OPEN/STOP” button to select program.
• Close the lid and turn lid knob to end position in clockwise direction.
• Press the “START” button.
• The rotor will start to turn and the red LED in the centrifuge lights up.
• At the end of the running time rotor break is applied until it comes to a complete standstill.
At the end of the running time, the red LED blinks until the lid is opened.
• The “UNLOCKED” display lights up; the interlocking mechanism is deactivated thereby
facilitating access to the rotor chamber. After expiry of the automatic unlocking time the
lid can be unlocked with idling rotor using the “OPEN/(EMERGENCY STOP)” button.
• Turn lid knob in an anticlockwise direction and open lid.
• Remove sample tubes.
• The centrifuge can be put into operation again immediately.
The centrifugation settings conform to the centrifugation recommendations for S-Monovettes
(see Brochure XX.780).
Concerning centrifugation conditions for other types of blood collection tubes refer to data
provided by the respective manufacturers.
Preset programs:
- Program 1 (Serum-Gel) 2,500 x g, 10 min.
- Program 2 (Urine) 500 x g, 5 min.
- Program 3 (variable), no preset