Instruction Manual
Centrifuge SMC 6 V
with swing-out rotor 6 x
SARSTEDT No.: 90.184.400
Страница 1: ...Instruction Manual Centrifuge SMC 6 V with swing out rotor 6 x SARSTEDT No 90 184 400...
Страница 2: ...bols 4 3 2 Device specific safety notes 4 3 3 Responsibilities of the owner 5 3 4 Responsibilities of operating personnel 5 4 Other information 6 4 1 Guarantee and liability 6 4 2 Shutdown disposal 6...
Страница 3: ...safe place for future reference Technical modification reserved N mbrecht June 2019 SARSTEDT AG Co KG Manufacturer Drucker Diagnostics 200 Shady Lane Suite 170 Philipsburg PA 16866 Phone 1 814 342 62...
Страница 4: ...explanatory information Non observance of these notes or information can lead to the device being operated incorrectly or even becoming damaged 3 2 Device specific safety notes This device has been bu...
Страница 5: ...or signs of damage Damaged power cord must not be used under any circumstances The device must only be connected to the mains voltage specified on the type label A damaged device must not be operated...
Страница 6: ...e warranty period the device may only be repaired by SARSTEDT AG Co KG or by persons authorized by SARSTEDT AG Co KG In case of improper use or repair the warranty will be void The device complies wit...
Страница 7: ...case of transport damage please contact the shipping agent immediately Observe the notification periods of the different freight carriers railway mail parcel service trucking company etc It can be as...
Страница 8: ...OPEN EMERGENCY STOP button Turn lid knob in counter clockwise direction and open lid There should be no foreign bodies in the rotor chamber Turn rotor by hand and check whether it runs freely and hor...
Страница 9: ...le red LED 10 START grey 11 OPEN STOP red By choosing the SMC 6 V you have decided in favour of a centrifuge that is very easy to use Three programs with different centrifugation speeds can be started...
Страница 10: ...lid knob to end position in clockwise direction Press the START button The rotor will start to turn and the red LED in the centrifuge lights up At the end of the running time rotor break is applied u...
Страница 11: ...preset 10 min Program 2 Urine Speed 1 805 rpm 90 min 1 RCF 500 x g 90 x g Time preset 5 min Program 3 variable Speed max 4 206 rpm RCF max 2 500 x g Time 0 to 30 min Power supply Electrical connectio...
Страница 12: ...Example for a disinfectant The following aqueous solution can be used as a disinfectant 25 g ethanol 96 35 g 1 propanol 0 1 g glyoxal ad 100 g aqua dest Apply the solution and allow it to react for a...
Страница 13: ...nt tubes causing further breakage If a tube breaks remove it carefully from the tube holder Dispose of the sample and tube fragments correctly and clean both the inside and outside of the tube holder...
Страница 14: ...and OPEN STOP buttons 3 sec until the LED display starts flashing 5 Release the buttons 6 Press the START button once per minute of the desired centrifugation time The time can be set within a range o...
Страница 15: ...ntrifuge SMC 6 V 10 Index Loading the rotor 10 Lid safety switch 6 15 Disinfectant 12 Machine data 11 Contamination questionnaire 16 Test run 8 Programming 14 Cleaning 12 Maintenance 13 Centrifugation...
Страница 16: ...Telephone No SN The instrument is free from dangerous substances The instrument came into contact with the following dangerous substances Groups of substances Substance name toxic caustic explosive ra...