2.4.3 Measurement tasks for the MiniTRACE CSDF
With the MiniTRACE CSDF you choose between the following operation modes:
Mode
Unit
Lid open
Lid closed
Dose rate H*(10) (Menu RATE)
Preferred direction observed, see Figure 27
µSv/h
-
X
Dose rate mean value H*(10) (Menu RMEAN)
Preferred direction observed, see Figure 27
µSv/h
-
X
Count rate (Menu CPS)
cps
X (
α
,
β
,
γ
)
X (
γ
)
Count rate mean value (Menu CMEAN)
cps
X (
α
,
β
,
γ
)
X (
γ
)
Activity (Menu BQ)
Bq
X (
α
,
β
,
γ
)
Not
recommended
Surface contamination (Menu BQCM2)
Bq/cm
2
X (
α
,
β
,
γ
)
Not
recommended
Count up (Menu COUNTUP)
cps
X (
α
,
β
,
γ
)
X (
γ
)
Food (Menu FOOD)
Bq/l
-
X (
γ
)
Table 2: Overview of measurements
Note
:
Measurement results in the “Surface contamination” (Bq/cm
2
) mode are based
on ISO 7503-1. For the activity mode (Bq) we have calibrated the instrument in
a 4
π
geometry.
The internal calibration factors are default values and not individually calibrated.
Figure 4: Opened protective lid
Protective
grid
Pancake Geiger-
Mueller detector
behind the grid
Lid open,
includes serial
number, infor-
mation and
energy filter