Coat-Endure
User Manual & Datasheet V 3.0
LED ERROR INDICATION
DESCRIPTION
TROUBLESHOOTING
RED-GREEN Blinking
Calibration Error
Recalibrate the instrument, make
sure that the probe is calibrated in
an empty metal-body tank.
RED Blinking
Probe Short-Circuit
Moisture deposition in the probe
connector. Clean the connector
and use the instrument.
GREEN Blinking
Probe Open
Remove the electronic insert from
the housing and check the cable
connections of the probe.
3 Times GREEN Blinking and 1
Red Blink
Illegal Key Combination
Switch all DIP switches to open
position. Use only legal combina-
tion of keys.
3 Times RED Blinking and 1
GREEN Blink
Circuit Error
Contact the Customer Support de-
partment at Sapcon.
Table 5: Error Indication