8.
Function mode
F01
Setting of High temp. alarm set temperature
F02
Setting of Low temp. alarm set temperature
F03
Display of battery accumulation time
F05
Setting of compressor delay time
F06
Service code input (code: 384)
F07 *
Temp. sensor Zero Adjustment
F08 *
Cascade sensor Zero Adjustment
F09
Factory test mode …… Unavailable
F10
Factory test mode …… Unavailable
F11
Factory test mode …… Unavailable
F12 *
Display of temperature in temp. sensor
F13 *
Display of temperature in cascade sensor
F14 *
Display of temperature in compressor sensor
F15 *
Display of temperature in AT sensor
F17 *
Model code setting (Initialization of non-volatile ROM and memory)
F18 *
Capillary heater is forcibly turned on/off
F19
Factory test mode …… Unavailable
F20
Factory test mode ... ... Unavailable
F21
Communication ID set ... ... Unavailable
F22
Communication mode set ... ... Unavailable
F23 *
Display of power supply voltage
F24
Remote alarm terminal output
F25
Alarm resume time set
F26 *
Display of actual operation rate
F27 *
Display of diagnosis value for running rate ... ... Unavailable
F28 *
Factory test mode …… Unavailable
F29 *
Factory test mode …… Unavailable
F30 *
ROM version is displayed
F31 *
Selectable buzzer ON/OFF setting in filter alarm
F32 *
Display of condensing fan motor accumulating time
Setting
In chamber temperature display, keep pressing
key over 5seconds to
procedure:
display “F00”.
Input the desired function code with
key and
key.
Press SET key to be function mode available.
* … You should input service code in F06 beforehand.
To cancel service code, input “000” in F06 or turn the power off.
F01:
<Purpose> To set high temp. alarm set temperature
<Operation> Input F01 and press SET key to display “010” (initial value).
Set the value in the range “005”~“020” with
key and
key.
Press SET key to store the value and revert to chamber temperature
display.
F02:
<Purpose> To set low temp. alarm set temperature
<Operation> Input F02 and press SET key to display “-10” (initial value).
Set the value in the range “-05”~“-20” with
key and
key.
Press SET key to store the value and revert to chamber temperature
display.
F03:
<Purpose> Battery accumulation time is displayed
<Operation> Input F03 and press SET key to display alternately F03 with “00.0”
(in case battery used for a month or less). Press SET key to revert to
chamber temperature display.
<Reset>
Input
‘
409
’ and press SET key to initialize battery accumulation time.
(’00.0’ is displayed)
18