F24:
<Purpose> Display of chamber temperature (digit in decimal point is displayed)
<Operation> In F24 displayed, press ENT key to display “000” (initial value)
Change the value to “001” with key and press ENT key.
Ex.) -85.1
㷄
㸢
851
Service code should be input in F06 prior to use this mode.
F25:
<Purpose> Setting of alarm resume time
<Operation> Input F25 and press ENT key to display “130” (initial value).
Change the value with key and key.
Press ENT key to store the value and revert to chamber temperature
display.
Alarm resume time (both buzzer and remote alarm relay are OFF by
pressing BUZZER key);
000: OFF 040: 40min. later
010: 10min. later 050: 50min. later
020: 20min. later 060: 60min. later
030: 30min. later
Alarm resume time (buzzer is OFF by pressing BUZZER key)
100: OFF 140: 40min. later
110: 10min. later 150: 50min. later
120: 20min. later 160: 60min. later
130: 30min. later (initial value)
F32:
<Purpose> Display of fan motor accumulation time
<Operation> Input F32 and press PV/SV key to display F32 and “000” (in case
accumulation time is a month or less) alternately.
<Cancel> Press PV/SV key to revert to chamber temperature display.
8.
Differential point (The point of compressor ON and OFF )
COMP
ON:
COMP OFF:
SV +0.5
㷄
or higher
SV -0.5
㷄
or lower
9.
Remote alarm
(1) For High/low temp. alarm (RLY 1)
In normal condition:
㸣
In alarm condition:
Remote alarm contact is N.O. N.C.
㸣
Remote alarm contact is N.C. N.O.
(2) For Power failure alarm (RLY 3)
In normal condition:
㸣
In power failure:
Remote alarm contact is N.O. N.C.
㸣
Remote alarm contact is N.C. N.O.