14
4. Function
mode
Setting range : 0~50
Display range : 0~59
Note) 04, 08, 13, 15, 16, 18~20, 26~29, 31, 33~49 are unused.
Setting process : In chamber temperature display, press key for 5 seconds to enter into
Function mode with ‘F00’ display. Change the 1
st
digit with key and
key.
Unacceptable
setting
range:
When you input a value in unacceptable setting range and press SET
key, error tone is emitted.
5. Error
code
E01: Temp. sensor is disconnected.
When temp. sensor is disconnected, ‘E01’ and ‘50’ are displayed
alternately.
Buzzer sounds and remote alarm activates.
Compressor is turned on.
Press BUZZER key to stop buzzer sounding.
Remote alarm keeps activating.
E02: Temp. sensor is short circuited.
When temp. sensor is short circuited, ‘E02’ and ‘-50’ are displayed
alternately.
Buzzer sounds and remote alarm activates.
Compressor is turned on.
Press BUZZER key to stop buzzer sounding.
Remote alarm keeps activating.
E09: Battery switch is OFF
When battery switch is off position or battery is unconnected, press
ALARM key to display ‘E09’.
6. Alarms
High temp. alarm : When chamber temperature is equal or higher than setting tempe
high temp. alarm setting tempe 1.0, ALARM lamp and the digital
display blink. After 10min. later, buzzer sounds and remote alarm output
activates.
When chamber temperature is equal or lower than setting temperature,
ALARM lamp and the digital display go off, buzzer stops sounding and
remote alarm output turns off.
If you press BUZZER key, buzzer stops sounding. Remote alarm keeps
activating.
High temp. alarm setting range: ‘005’ ~ ‘015’
Default setting: ‘010’
Low temp. alarm :
When chamber temperature is equal or lower than setting temperature -
low temp. alarm setting temperature - 1.0, ALARM lamp and the digital
display blink. After 10min. later, buzzer sounds and remote alarm output
activates.
When chamber temperature is equal or higher than setting temperature,
ALARM lamp and the digital display go off, buzzer stops sounding and
remote alarm output turns off.
If you press BUZZER key, buzzer stops sounding. Remote alarm keeps
activating.
Low temp. alarm setting range: ‘-05’ ~ ‘-15’
Default setting: ‘-10’