Measurement of
Proper range to be used
250-1000 volts
1000V
250V
50V
10V
2.5V
50-250 volts
10-50 volts
2.5-10 volts
0-2.5 volts
c Test Lead Connections.
See GENERAL INST
RUCTIONS^.
The test leads connected to the meter are applied
across the voltage to be checked. Since alternating
current is being measured, the readings are correct
to which side of the voltage the test leads are con
nected. (Fig. 2-3 & -4)
d Scale Reading.
Use the red scale third from the
top marked
AC 1 0 V UP
for all AC voltage measure
ments except for the 2.5 volt range, for which the
black scale below marked
AC 2 .5 V
is used. (Fig. 2-5)
13 —