Sanwa SE300 Скачать руководство пользователя страница 2

DRAWING No. SE300 01-1408 2040 6006

5-4  Contact measurement (Using the optional ENC-3)

1. Do not perform measurement if the contact adapter, contact 

maker or rim speed ring of the contact measurement attachment 

is damaged, worn or deformed.

2. Be sure to tighten the contact adapter fixing screw tightly.

3. Fit the contact maker or rim speed ring all the way into the 

root of the rotary shaft of the contact adapter.

4. When measuring a high-speed rotating object (8000 rpm or 

higher), do not measure it continuously for more than one minute.

5. Do not use a tripod in the contact measurement.

6. Do not use the contact maker or rim speed ring with a high-

temperature rotating object.

WARNING

The contact measurement required the optional ENC-3 Contact 

Measurement Attachment.

1) Attach the contact adapter as shown below and secure the fixing 

screw tightly.

Contact adapter

Fixing screw

2) Select the contact maker or rim speed ring according to the 

purpose of measurement and fit it on the tip of the contact 

adapter.

                 

Contact maker

Rim speed ring

* Select the rim speed ring when measuring the velocity of a belt 

conveyer or similar object (Travel distance per minute in m/min or 

travel distance per second in m/s).

3) Turn the instrument ON and select the contact measurement 

mode (see “4-3 Measurement mode selection”). Ensure that 
the

indicator appears on the display panel and set the desired 

measurement mode.

Each press of the 

 button switches the measurement 

mode in the following cycle:

Rotation speed measurement   Rotation speed measurement   Interval measurement

(revolutions per minute) 

(revolutions per second)  (milliseconds)

 

 

Rim speed measurement      Rim speed measurement      Rotation count measurement  

 

(meters per second)      (meters per minute)        (counts)

When the rim speed ring is attached, select either the [m/min] or 

[m/s] mode.

4) Apply the contact maker or rim speed ring on the rotating object. 

Special care is required for the method of application.

The contact maker should be applied perpendicularly to the 

contact surface toward the axis of the rotating object.

The rim speed ring should be applied in parallel with the 

measurement target.

Rotating object

     

5) The instrument displays the measurement result.

 

6) Detach the contact adapter by loosening the fixing screw, 

position the instrument so that the side with the fixing screw 

faces down and pull the adapter obliquely considering the 

presence of a lock mechanism.

    

5-5  Measurement modes

• [rpm] 

: Revolutions per minute

• [rps] 

: Revolutions per second

• [ms] 

: Reflected light detection interval (milliseconds)

• [count]  : Rotation count measurement

• [m/min]  : Travel distance per minute (meters)

Rotation speed(rpm)xCircumference of rim speed ring 100(mm)

1000

• [m/s] Travel distance per second (meters)

Rotation speed(rps)xCircumference of rim speed ring 100(mm)

1000

[6] MAINTENANCE

1. The information in this section is critical for safety. Manage 

the instrument based on full understanding of the information 

given in this manual.

2. For safety and accuracy of the instrument, calibrate and 

inspect it at least once a year.

WARNING

6-1  Maintenance and Inspection

Be sure to perform the maintenance inspection of the instrument to 

use it safely for an extended period and keep the high quality. The 

inspection items are as described in “5-1 Start-up inspection”.

6-2   Calibration and Inspection

For more information, please contact Sanwa’s authorized agent /    

distribute service provider, listed in our website.

See section 7-3.

6-3  Battery replacement

• Batteries shipped with the instrument

The batteries provided with the instrument are monitoring batteries. 

Their service life may be shorter than brand-new batteries.

* Monitoring batteries refer to the batteries for use in checking the 

functionality and performance of the product.

                           

1

 Using a Phillips screwdriver, loosen the retaining screw of the 

battery cover.

2

 When the battery cover comes up, remove it.

3

 Replace both of the two batteries in the battery holder with new 

ones by taking care of the polarity.

4

 Attach the battery cover and secure the retaining screw.

6-4  Cleaning and Storage

1. The instrument is not resistant to volatile solvents and should not 

be wiped with lacquer thinner or alcohol. When it gets dirty, wi pe 

dirt away with a soft cloth moistened with a small amount of water.

2. The instrument is not resistant to heat and should not be 

placed near a source of high heat.

3. Do not store the instrument in a position subjected to excessive 

vibrations or a position presenting the risk of dropping.

4.  Avoid storing the instrument under direct sunlight, high or low 

temperature, high humidity or a position with a risk of condensation.

5. Be sure to remove the batteries when the instrument is 

expected to be left unused for a long period of time.

CAUTION

6-5  Storage in the carrying case

The instrument can be stored in the carrying case as shown below.

Main instrument

Contact maker (*)

Contact adapter (*)

Test certificate

Reflective stickers

Rim speed ring (*)

Batteries

Instruction manual

(*) Optional ENC-3 Contact Measurement Attachment

[7]  AFTER-SALE SERVICE

7-1  Warranty and Provision

Sanwa offers comprehensive warranty services to its end-users 

and to its product resellers. Under Sanwa's general warranty 

policy, each instrument is warranted to be free from defects in 

workmanship or material under normal use for the period of one (1) 

year from the date of purchase.

This warranty policy is valid within the country of purchase only, 

and applied only to the product purchased from Sanwa authorized 

agent or distributor.

Sanwa reserves the right to inspect all warranty claims to 

determine the extent to which the warranty policy shall apply. This 

warranty shall not apply to disposables batteries, or any product or 

parts, which have been subject to one of the following causes:

1. A failure due to improper handling or use that deviates from the 

instruction manual.

2. A failure due to inadequate repair or modification by people 

other than Sanwa service personnel.

3. A failure due to causes not attributable to this product such as 

fire, flood and other natural disaster.

4. Non-operation due to a discharged battery.

5. A failure or damage due to transportation, relocation or dropping 

after the purchase.

7-2  Repair

Customers are asked to provide the following information when 

requesting services: 

1. Customer name, address, and contact information

2. Description of problem

3. Description of product configuration

4. Model Number

5. Product Serial Number

6. Proof of Date-of-Purchase

7. Where you purchased the product

Please contact Sanwa authorized agent / distributor / service 

provider, listed in our website, in your country with above 

information. An instrument sent to Sanwa / agent / distributor 

without above information will be returned to the customer.

Note:

1) Prior to requesting repair, please check the following:

Capacity of the built-in battery, polarity of installation and 

discontinuity of the test leads.

2) Repair during the warranty period:

The failed meter will be repaired in accordance with the 

conditions stipulated in 7-1 Warranty and Provision.

3) Repair after the warranty period has expired:

• If it is expected that servicing can restore the original functioning of 

the instrument, we will service it for a price upon request of the user.

• The service charge or transport freight could sometimes become higher 

than the product price. Please consult us before asking for servicing.

• The minimum retention period of the servicing performance parts of 

this instrument is six (6) years after the discontinuation of production. 

This period is equal to the servicing available period. However, the 

retention period of a part may be reduced if it becomes unavailable 

due to discontinuation of production of the part manufacturer, etc.

4) Precautions when sending the product to be repaired:

To ensure the safety of the product during transportation, place the 

product in a box that is larger than the product 5 times or more in 

volume and fill cushion materials fully and then clearly mark 

“Repair Product Enclosed” on the box surface.  The cost of 

sending and returning the product shall be borne by the customer.

7-3  SANWA web site

http://www.sanwa-meter.co.jp

E-mail: [email protected]

[8] SPECIFICATIONS
8-1  General Specifications

Measurement methods

Red visible light photoelectric reflection method, 
contact method (*)99999

Measurement modes

rpm, rps (Revolution speed)
ms (Interval)
count (Rotation count)
m/min, m/s (Rim speed) (*)

Reflection detection 
distance

Approx. 50 to 500 mm

Display panel

LCD, 5 digits, backlit

Display updating intervals

Approx. 1 sec. (rpm, rps, cycle, m/min, m/s modes)
Approx. 0.1 sec. (count mode)

Operating environmental 
conditions

Altitude ≤ 2000 m, indoor, pollution degree II

Operating temperature/
humidity range

5 °C to 40 °C, ≤80 %RH (without condensation)

Storage temperature/
humidity range

-10 °C to +40 °C, ≤80 %RH (without condensation)
(The batteries should be removed before long-
term storage.)

Auto power OFF

The instrument turns OFF in about 2 minutes 
after the last operation or last reflected light 
detection.

Over-scale display

“OL” is displayed on the display panel.

Low battery alarm

When the battery power is consumed and the 
voltage drops below about 2.3 V, the 

indicator appears on the display panel

Power supply

“AA”-size dry cell battery (R6P/LR6) x 2

Battery life

Approx. 28 hours (alkaline batteries, backlight 
not used)

Dimensions

210(

H

) x 60(

W

) x 50(

D

) mm (Main instrument only)

298(

H

) x 60(

W

) x 50(

D

) mm (Contact a 

Contact maker)

Weight (incl. batteries)

Approx. 218 g(Main instrument only), 
Approx. 278 g(Contact a Contact maker)

Accessories

Instruction Manual, “AA”-size dry cell battery 
(R6P/LR6) x 2, carrying case, reflective stickers 
(x 50)

8-2  Optional accessories

ENC-3

Contact Measurement Attachment (Contact adapter, contact 
maker and rim speed ring)

                                  

     

SE-A30

Contact maker

                                       

SE-A31

Rim speed ring, circumference 100 mm

                                      

SE-T3

Reflective stickers (x 100 = 50-sticker sheet x 2)

C-SE300

Carrying case

8-3  Measuring ranges and accuracies

Accuracy guaranteed temperature/humidity ranges:23±5 °C, no more 

than 80 %RH, (without condensation.)

rdg: Reading. dgt: Lowest digits.
Measurement ranges

Measurement mode

Noncontact 

measurement

Contact

measurement

rpm

30.0 – 99999

30.0 – 19999

rps

0.50 – 1600.0

0.50 – 333.00

ms

0.600 – 1999.0

3.000 – 1999.0

count

0 – 99999

0 – 99999

m/min

 – 

3.0 – 1999.0

m/s

 – 

0.05 – 33.00

(*) When the optional ENC-3 is used

Accuracy     ±(0.03 % rdg + 1dgt)

*  Accuracy in noncontact measurement. It does not include errors 

that may be caused by extraneous light or shake of the instrument.

* In the contact measurement, the accuracies and slipping movements of 

the contact maker and rim speed ring are added to this figure above.

* In the rim speed measurement, ±0.5 % rdg is added to the figure above.
Accuracy calculation method

Example) Noncontact revolution speed measurement (rpm)

Displayed value: 10000 rpm

Accuracy: Error: ±(10000 rpm x 0.03 % + 1 dgt) = ±4 rpm

True value: Within the range of 10000 rpm ±4 rpm (9996 – 10004 rpm)

* AT 10000 rpm, 1 dgt corresponds to 1 rpm.
For the purpose of improvement, the design and specifications of the 

products described above may be subject to change without notice.

(*) When the optional ENC-3 is used

−  5  −

−  6  −

−  7  −

−  8  −