4-5 Range Hold
Press the RANGE button momentarily to set the manual range
mode then 'AUTO' disappears in the display. In manual range
mode, press the button again to step through the ranges. To
return to the auto mode, press the button for 1 sec. or more then
AUTO is shown.
Note:
Manual mode is not available in Hz measurement.
4-6 Data Hold
When the HOLD button is pressed, the data display at that time is
hold ('H' is shown on the display). The display will not changed
while the function is active. Press the button again to cancel the
function. ('H' on the display disappears.)
4-7 Auto Lead Resistance Calibration
When entering the 50
Ω
range manually by RANGE button for
high precision low resistance measurement, this feature will
prompt you to short the inputs for calibration. The display shows
"Shrt". Simply short the leads for about 3 seconds until the display
shows zero, then the resistance in the leads and in the internal
protection circuitry of the meter is compensated automatically.
The compensation value can be as much as 5
Ω
. PC510:If you
need a compensation value that is higher than that, Relative
mode is recommended.