Sanwa MG1000 Скачать руководство пользователя страница 2

4-2-2  Voltage measurement function (AC/DCV)

Range

Resolution

Accuracy

Input Resistance Max. Overload Protection

AC/DC 600V

1V

±

(3%rdg+2dgt)

9M

780Vrms

Note
ACV and DCV are automatically detected.
Sensitivity: 2 V or higher.
AC frequency bandwidth: 45-400 Hz
The ALAMP indicator lights up when the input is 

30 V AC/DC.

“      ” appears and buzzer sounds if the input is 

600 Vrms.

Maximum display value: 780 V.

4-2-3  4000 

 measurement function / Continuity buzzer ( 

 )

Range

Resolution

Accuracy

Input Resistance Max. Overload Protection

4000 

1

±

(3%rdg+3dgt)

3V

600Vrms

Note
A buzzer sounds below about 40 

.

A buzzer sounds and the ALARM indicator lights up when the input is 

30 V 

AC/DC.

4-2-4  40 

 measurement function ( 

 )

Display Range

40.00 

Measurement range

0.01 

 to 40.00 

Accuracy

±

(3% + 10 dgt) (after 0 

 ADJ)

Resolution

0.01 

Open Circuit voltage

6V (typical value)

Short-circuit current

200 mA minimum (

 2.0 

)

Protection

Fuse (0.5 A/600 V)

Note
The bargraph display is disabled.

[5]  MEASURING PROCEDURE

5-1  Opening/closing the body cover

The body cover cannot be closed unless 
the function switch is OFF.

5-2  Start-up inspection

5-3  Voltage measurement method

Note
ACV and DCV are automatically 
detected.

5-4  Insulation resistance measurement method

5-4-1  Buttons used in the measurement

Measurement: 

MEASURE

 button, 

0

ADJ

 button (to set the max test voltage function)

Continuous measurement: 

LOCK

 and 

MEASURE

 buttons.

5-4-2  Measurement method

1. Connect the test leads to the object (circuit) to be measured.

If the object to be measured is a live circuit (

30 V AC/DC), the voltage is 

displayed, buzzer sounds, and the ALARM indicator lights up. In this case, the 
test voltage cannot be output by pressing the MEASURE or LOCK button. 
Disconnect the power supply from the object to be measured and try again.

2. Output the test voltage.

    Start measurement (

MEASURE

 button)         Continuous measurement (

LOCK

 button)

3. End the measurement.                           3. End of continuous measurement

At the moment the test voltage output is stopped, the last displayed value is held 
automatically (      indication) and the measured object (circuit) is discharged 
automatically.
When the measured object (circuit) is discharged below 30 V, the     indicator 
goes off.

Note

1. An oscillation tone may be heard during measurement, but this is not malfunction.
2. The automatically held value is retained until the start of the next measurement 

or the tester enters APS mode. The held value is released and the new 
measurement value is displayed when the next measurement is started.

3.  If a measuring point of certain equipment for insulation resistance measurement is 

unknown, contact the manufacturer of that equipment to ensure correct measurement.

5-5  4000 

 Resistance measurement / Continuity check method

5-5-1  Buttons used in measurement

No button is used.

5-5-2  Measurement method

1. Set the power/function switch to “4000 

.”

2. Connect the test lead to the object to be measured.
3. Read the displayed value.

5-6  40 

 measurement method

5-6-1  Buttons used in measurement

Measurement: 

MEASURE

 button and 

0

 ADJ

 button.

Continuous measurement: 

LOCK

 button, 

0

 ADJ

 and 

MEASURE

 button.

5-6-2  Measurement methods

Method using the 

MEASURE

 button:

1.  Set the function switch to “40.00 

.”

2. Short-circuit the test leads, press the MEASURE button, and release the 

MEASURE button when the tester detects a stable reading. The displayed 
value will be held.

3. Press the 0

 ADJ button to reset the displayed value to “0.00 

.” [0

 ADJ] 

should light up at this time.

4. Connect the test leads to the object to be measured, and press the 

MEASURE button.

5. To end measurement, release the MEASURE button. The last displayed 

value will be held automatically.

Method using the 

LOCK

 button:

1.  Set the function switch to “40.00 

.”

2.  Short-circuit the test leads and press the LOCK button.
3. When the tester detects a stable reading, press the 0

 ADJ button to reset 

the displayed value to “0.00 

.” [0

 ADJ] should light up at this time.

4.  Connect the test leads to the object to be measured and read the displayed value.
5. To end measurement, press the MEASURE or LOCK button. The last 

displayed value will be held automatically.

Caution: The measurement consumes a large amount of current. Use caution 

with the battery exhaustion.
The displayed value may not become “0.00” depending on the 
contact condition while shorting the test leads.

[6]  MAINTENANCE

6-1  Maintenance inspection

1) Appearance: Is the meter not damaged due to falling or other cause?
2) Test leads: Are the test leads not damaged, or is the core wire not exposed 

from any part or broken?

If any of the above problems exists, stop using the meter and request for repair.

6-2  Calibration / Inspection

For more information, please contact your dealer or Sanwa agent.

6-3  Storage

Check if the instrument body appears damaged.

Check if      appears on the top right of the LCD. If 

 

appears or blinking, replace all batteries with new ones (see 
section 6-4-1).

Select the 4000 

  measurement function, short the test lead 

and alligator clip, and ensure the displayed value is no more 
than 1 

. If “OL” is displayed, replace the test leads as they may 

be damaged.

Select the 40 

 measurement function, short the test lead and 

alligator clip, and ensure the displayed value is no more than 
0.05 

.  If “OL” is displayed, replace the fuse as it may be 

damaged (see section 6-4-2).

Inspection/ Check

Appearance

Internal batteries

Test leads

Fuse

Method

The test voltage is output when the 

MEASURE

 

button is held pressed.        appears and short 
buzzer sounds during this operation.

Press and hold the 

LOCK

 button for more than 

2 seconds to output the test voltage 
continuously.        and            appears and short 
buzzer sounds during this operation.

6-4  Battery / Fuse replacement

Battery when the meter is shipped: 
A battery for monitoring has been installed prior to shipment from the factory.  It 
may be discharged before the expiration of the described battery life.
*The battery for monitoring is a battery used to check the functions and 

performance of the product.

6-4-1  Battery replacement: R6 1.5V x 6pcs (MG125, MG500) ; LR6 1.5V x 6pcs (MG1000)

Note: Use caution with battery polarity when replacing batteries.

6-4-2  Fuse replacement

Fuse: Part No. F1202, Ø6 X 30 mm, 0.5 A/600 V, breaking capacity 20 kA.

[7]  AFTER-SALE SERVICE

7-1  Warranty and Provision

Sanwa offers comprehensive warranty services to its end-users and to its product 
resellers. Under Sanwa's general warranty policy, each instrument is warranted to 
be free from defects in workmanship or material under normal use for the period of 
one (1) year from the date of purchase.
This warranty policy is valid within the country of purchase only, and applied only to 
the product purchased from Sanwa authorized agent or distributor.
Sanwa reserves the right to inspect all warranty claims to determine the extent to which 
the warranty policy shall apply. This warranty shall not apply to disposables batteries, or 
any product or parts, which have been subject to one of the following causes:
1. A failure due to improper handling or use that deviates from the instruction 

manual.

2. A failure due to inadequate repair or modification by people other than Sanwa 

service personnel.

3. A failure due to causes not attributable to this product such as fire, flood and 

other natural disaster.

4. Non-operation due to a discharged battery. 
5. A failure or damage due to transportation, relocation or dropping after the 

purchase.

7-2  Repair

Customers are asked to provide the following information when requesting services:
1. Customer name, address, and contact information
2. Description of problem
3. Description of product configuration
4. Model Number
5. Product Serial Number
6. Proof of Date-of-Purchase
7. Where you purchased the product
Please contact Sanwa authorized agent / distributor / service provider, listed in 
our website, in your country with above information. An instrument sent to Sanwa 
/ agent / distributor without above information will be returned to the customer.

Note:
1) Prior to requesting repair, please check the following:

Capacity of the built-in battery, polarity of installation and discontinuity of the test leads.

2) Repair during the warranty period:

The failed meter will be repaired in accordance with the conditions stipulated in 
7-1 Warranty and Provision.

3) Repair after the warranty period has expired:

In some cases, repair and transportation cost may become higher than the price of 
the product. Please contact Sanwa authorized agent / service provider in advance.
The minimum retention period of service functional parts is 6 years after the discontinuation 
of manufacture. This retention period is the repair warranty period. Please note, however, if 
such functional parts become unavailable for reasons of discontinuation of manufacture, 
etc., the retention period may become shorter accordingly.

4) Precautions when sending the product to be repaired To ensure the safety of the 

product during transportation, place the product in a box that is larger than the 
product 5 times or more in volume and fill cushion materials fully and then clearly 
mark “Repair Product Enclosed” on the box surface. The cost of sending and 
returning the product shall be borne by the customer.

7-3  SANWA web site

http://www.sanwa-meter.co.jp/
E-mail: [email protected]

Release the 

MEASURE

 button to stop 

outputting the test voltage.

Press the 

MEASURE

 or 

LOCK

 button to 

stop outputting the test voltage.

     WARNING

1. Be sure to conduct the start-up inspection described in section 5-2 below.
2. Before measurement, disconnect power supply from the object (circuit) to 

be measured.

3. After measurement, be sure to discharge the high voltage charged in the 

measured object.

4. During measurement, keep your fingers behind the finger guards of the test 

leads and do not touch the metallic part of the alligator clips.

5. After measurement, be sure to turn the power/function switch to “OFF” 

position to prevent electric shock and battery exhaustion.

     CAUTION

1. When selecting the maximum test voltage (insulation resistance 

measurement), hold the 0 _ ADJ button pressed while turning the 
power/function switch. Otherwise, the display shows “Err0” and the test 
voltage will not be output even if the MEASURE or LOCK button is pressed.

2. When the object to be measured is grounded, usually the alligator clip is 

connected to the ground (EARTH) and the test lead to the measured circuit 
(LINE). (This connection method usually results in a lower measurement 
value than the opposite connection method.)

3. To prevent measurement errors, prevent the test lead connected to the 

LINE terminal from touching the object to be measured or ground whenever 
possible.

4. The insulation resistance varied greatly depending on the temperature and 

humidity as well as on the output voltage (test voltage). In general, higher 
temperature, humidity and/or output voltage mean lower insulation 
resistance.

5. Measurement is inhibited if voltage (

30 V AC/DC) is present in the circuit.

6. Keep in mind that the time of auto discharge will be delayed if fuse has 

broken.

     WARNING

1. The following instructions are very important for safety. Read this manual 

thoroughly to ensure correct maintenance.

2. Calibrate and inspect the meter at least once a year to ensure safety and 

maintain its accuracy.

     CAUTION

1. The panel and case are not resistant to volatile solvent and must not be 

cleaned with thinner or alcohol.

2. The panel and case are not resistant to heat.  Do not place the tester near 

heat-generating devices.

3. Do not store the tester in a place where it may be subjected to vibration or 

from where it may fall.

4. Do not store the tester in places under direct sunlight, or hot, cold or humid 

places or places where condensation is anticipated.

5. If the tester will not be used for a long time, remove the battery.

     WARNING

1. If the rear case is removed with an input being applied to the measuring 

terminals, you may suffer electric shock.  Before starting replacement, 
always make sure no input is being applied and the function switch is OFF.

2. Be sure to use the fuse of the specified rating and specification. Never use 

a substitute nor short the fuse terminals.

Spare fuse
(optionally available)

Short beeps

GND

In case of Live circuit

OFF

Green
indication
ON/blinking

Red indicator ON

Hold for
more than
2 sec. and
then release.

Beep

Grounding

side

(EARTH)

Line side

(LINE)

Release

Red 
indicator 
ON

Buttons used

MΩ 

MΩ 

MΩ 

MΩ 

MΩ 

MΩ 

Short beeps

Green
indication
ON/blinking

Press

Press here 

or 

here.

No.0436-002AB