- 20 -
- 19 -
2) Measuring ranges
Function Measuring
range
Accuracy
Remarks
Resistance
measurement
Continuity
Check
999.9 Ω
99.99 kΩ
999.9 kΩ
±(1.5 %rdg + 7 dgt)
≦
30 Ω: Beeps
9.999 kΩ:
Not supported
Over 1 MΩ: "OL"
①
Set
function. (Ω or kΩ will be indicated.)
②
Connect the alligator clip to the circuit to be measured
and the measurement probe too.
③
Read the display.
Note:
• MΩ measurement button is not available.
•
Open circuit voltage between the input terminals:
Approx. 2.0
~
2.5 Vdc
1. The followings are important to safety. Read this
manual throughly to maintain the instrument.
2. Calibrate and inspect the instrument at least once a
year to ensure safety and maintain its accuracy.
1. Do not open the battery door with live input terminals
to avoid electric shock.
2. Make sure the power is OFF, before starting the
replacement.
[6] MAINTENANCE
WARNING
6-1 Simple Examination
1) Appearance:
•
Check that the instrument does not look damaged
caused by dropping etc.
2) Measurement Probe, Alligator Clip, Earth lead:
•
Check that the measurement probe and the alligator clip
fit on the instrument.
•
Check for the earth lead especially exposed core wire
anywhere on it.
If you find any problem on the above items, stop using
immediately and ask us to repair it.
6-2 Calibration
For requesting calibration and inspection, contact an
authorized agent/distribution service provider, listed in our
website. (See 7-2 section 4)
6-3 Battery Replacement
WARNING
Pre-installed battery
Since the pre-installed battery is for monitoring, it may not
be durable as typically expected.
*The purpose of the battery for monitoring is to check for
the functions and performances of the product.
①
②
③
①
②
③