Sanwa DCL11R Скачать руководство пользователя страница 2

 ①

 Using a Phillips screwdriver, remove the 

fixing screw (x 1) from the battery 

compartment cover.

 ②

 Remove the battery compartment cover.

 ③ 

Replace both of the batteries in the 

battery holder with new ones by 

observing the correct polarity.

 ④ 

Place the battery compartment cover 

and tighten the fixing screw in the 

original position.

[6]  MAINTENANCE

1. This section is very important for safety. Read and understand the following 

instructions fully and maintain your instrument properly.

2. To maintain safety and accuracy, calibrate and inspect the instrument at 

least every year.

6-1   Maintenance and Inspection

Before use, check the instrument to confirm that there is no abnormality and that 

the CT can be opened and closed smoothly. If anything is abnormal, do not use the 

instrument and return it to your authorized Sanwa agent or distributor for repair.

6-2   Calibration

For more information, please contact Sanwa’s authorized agent / distribute 

service provider, listed in our website.

 

See section 7-3.

6-3  Cleaning and Storage

1. The panel and case are not resistant to volatile solvent and must not be 

cleaned with thinner or alcohol. If the instrument gets dirty, wipe with a soft 

cloth moistened with a small amount of water.

2. The panel and case are not resistant to heat. Do not place the instrument 

near heat-generating devices (such as a soldering iron).

3. Do not store the instrument anywhere it may be subject to vibrations or could fall.

4. When storing the instrument, avoid hot, cold or humid locations, locations 

exposed to direct sunlight, or locations where condensation is anticipated.

5. When the instrument is not going to be used for an extended time, be sure 

to remove the batteries.

6-4  Battery replacement

• Batteries shipped with the instrument

The batteries loaded at the factory are monitor batteries, so their service life 

may be shorter than that of brand-new batteries. Monitor batteries are a type 

of battery used to check the functions and performance of a product.

1.  To avoid electric shock, do not remove the battery compartment cover while an 

input is applied to the clamp sensor or the instrument is measuring something.

2.  Make sure that the instrument is set to OFF before replacing the batteries.

[7]  AFTER-SALE SERVICE
7-1  Warranty and provision

Sanwa offers comprehensive warranty services to its end-users and to its 

product resellers. Under Sanwa's general warranty policy, each instrument is 

warranted to be free from defects in workmanship or material under normal use 

for the period of one (1) year from the date of purchase.

This warranty policy is valid within the country of purchase only, and applied 

only to the product purchased from Sanwa authorized agent or distributor.

Sanwa reserves the right to inspect all warranty claims to determine the extent 

to which the warranty policy shall apply. This warranty shall not apply to 

disposables batteries, or any product or parts, which have been subject to one 

of the following causes:
1. A failure due to improper handling or use that deviates from the instruction manual.

2. A failure due to inadequate repair or modification by people other than Sanwa 

service personnel.

3. A failure due to causes not attributable to this product such as fire, flood and 

other natural disaster.

4. Non-operation due to a discharged battery.

5. A failure or damage due to transportation, relocation or dropping after the 

purchase.

7-2  Repair

Customers are asked to provide the following information when requesting 

services: 
1. Customer name, address, and contact information

2. Description of the problem

3. Description of the product configuration

4. Model Number

5. Product Serial Number

6. Proof of Date-of-Purchase

7. Where you purchased the product

 

Please contact Sanwa authorized agent / distributor / service provider, listed 

in our website, in your country with above information. An instrument sent to 

Sanwa / agent / distributor without above information will be returned to the 

customer.

Note:

1) Prior to requesting repair, please check the following:

Capacity of the built-in battery, polarity of installation and discontinuity of the 

test leads.

2) Repair during the warranty period:

The failed instrument will be repaired in accordance with the conditions 

stipulated in 7-1 Warranty and provision.

3) Repair after the warranty period has expired:

• If it is expected that servicing can restore the original functioning of the product, 

we will service it for a price upon request of the user.

WARNING

CAUTION

WARNING

•  The service charge or transport freight could sometimes become higher than the 

product price. Please consult us before asking for servicing.

• The minimum retention period of the servicing performance parts of this 

product is six (6) years after the discontinuation of production. This period is 

equal to the servicing available period. However, the retention period of a part 

may be reduced if it becomes unavailable due to discontinuation of production 

of the part manufacturer, etc.

4) Precautions when sending the product to be repaired:

To ensure the safety of the product during transportation, place the product in 

a box that is larger than the product 5 times or more in volume and fill cushion 

materials fully and then clearly mark “Repair Product Enclosed” on the box 

surface.  The cost of sending and returning the product shall be borne by the 

customer.

7-3  SANWA web site

http://www.sanwa-meter.co.jp

E-mail: [email protected]

[8]  SPECIFICATIONS
8-1  General Specifications

Operation method

Δ -

Σ

 method

AC detection method

True RMS method (AC coupling)

Display

6000 count

Sampling rate

Approx. 2 times/sec

Overload display

“OL” on the numeric display

Range selection

Auto range

Low battery indication

When the batteries are discharged down to 2.3 V or 

less, the 

 indicator lights on the display.

Current measurement 

method / aperture 

diameter

Clamping current sensor (CT) /

Max. approx. dia. 

φ

22 mm

Operation 

environment

Altitude max. 2000 m, indoor use, environmental 

pollution degree II. 

Operating 

temperature/ 

humidity 

Temperature 5 to 40 °C, 

Relative humidity max. 80 % RH at 5-31 °C and 

decreases linearly through 80 % RH to 50 % RH at 

31-40 °C (without condensation)

Storage temperature/

humidity 

-10 to 40 °C: Max. 80 % RH (without condensation)

40 to 50 °C: Max. 70 % RH (without condensation)

Power supply

“AAA”-size alkaline battery LR03 x 2

Auto power OFF

Power off in approx. 15 min. after last operation

Power consumption / 

Battery life

Approx. 1.9 mA / Approx. 500 hours (backlight 

extinguished)

Dimensions and mass

145  (

H

) x 54 (

W

) x 31 (

D

) mm, approx. 120 g 

(including batteries) 

Safety standards

IEC61010-1, IEC61010-2-030 CAT. III 300 V, 

IEC61010-2-32

EMC directive

IEC61326-1 

Accessories

Instruction manual, carrying pouch (C-DCL10)

8-2 Measuring ranges and accuracies

Accuracy-guaranteed temperature/humidity range: 23 ±5 °C, max. 80 % RH 

(without condensation).

rdg: Reading 

dgt: digits (lowest digit)

Accuracy-guaranteed range: 1 % to 100 % of measurement range

Crest factor CF: Full-scale CF < 1.6, half scale CF < 3.2

ACA measurement RMS response, AC coupling (sine wave alternate current)

Range

Resolution

Accuracy: 50/60 Hz

Accuracy: 45-400 Hz

60.00 A

0.01 A

±(2.0 %rdg + 5 dgt)

±(3.0 %rdg + 10 dgt)

300.0 A

0.1 A

Accuracy measurement method

(Example)

 

ACA measurement

Reading: 10.00 A

Range accuracy: 60.00 A range = ±(2 %rdg + 5 dgt)

Error: ±(10.00 A x 2.0 % + 0.01 A + 5) = ±0.25 A

True value: 10.00 A ±0.25 A (in the range from 9.75 to 10.25 A)

* 3 dgt in the 60.00 A range is equivalent to 0.03 A.

* 3 dgt in the 300.00 A range is equivalent to 0.3 A.

Specifications and external appearance of the product described above 

may be revised or modified without prior notice.

−  3  −