42/52
The charge status is displayed as follows by a
red
LED:
•
If the LED is lit continuous red the battery pack is being charged quickly in the
first stage. The battery pack is between 30 and 80% full.
•
If the LED is flashing slowly then the charger has switched to trickle charging
and the battery pack is between 80 and 100 % full.
•
If the LED flashes very rapidly then it indicates that the battery pack has been
discharged too far and is almost completely empty. Normally a "healthy" battery
pack can be revived by the charger unit and after a certain time it will switch
over to quick charging and the LED goes on continuously. Completely
discharging the battery should be avoided, since the service life of the batter
pack will be greatly reduced.
•
As a rule "deep" discharges only happen if the battery pack is not recharged
regularly between uses and the battery has no time to recover.
8.1.1 Technical data
Mains voltage supply (50/60 Hz, +/-15 %)
100-240 V AC
Nominal rating (input)
36 W
Charging voltage
24 V DC
Theoretical charging current
1.0 A
Protection class
IP40
Safety approval
GS and UL
8.1.2 Safety guidelines
•
Protect against damp
•
Only charge in a well ventilated room
•
Do not pull the plug out of the socket by its cord
8.1.3 Use only for the purpose intended
-
The charger unit is designed exclusively for charging
AGM cell batteries
containing liquid, gel-type and fabric-type electrolyte (
such as the battery pack
supplied with the Liftkar PT).
- It is not permissible to charge NiCd or NiMH batteries or primary cells.